46
MBW 473 Manual English V5.0
At equilibrium, the water vapor pressure is assumed to be ho-
mogenous, but the user should determine this by experimenta-
tion according to the type of chamber being tested. Temperature
probes should be positoned according to the need for validation.
The most common arrangement is four probes in the top
corners, four in the bottom corners and a single probe in the
center. For larger chambers, more probes can be added so that
the entire working volume is measured. This allows for the full
spatial volume to be temperature mapped. By combining this
temperature data with the measured dew point, %RH can also
be mapped. Software such as that supplied with the MBW T12
temperature system can be used to display and record the dew point, temperature and calculated
%RH values. Please contact MBW or your local supplier for more information.
RH Generator Validation
Transportable RH Generators provide a convenient and fast
way of calibrating RH instruments. Most use RH sensor con-
trol probes to measure and control the conditions in the gen-
erator
’s test chamber. The standard calibration uncertainty of
the RH generator is limited by calibration drift, linearity and
temperature coefficients of the control probe. With the use of
a dew point mirror, with its inherent stability and precision,
the uncertainty can be significantly improved.
The 473-RP2 is ideal for this application. As shown, the RP2
measuring head is introduced into the generator chamber
through a front access port. The measuring cell of the mirror
is exposed to the moving air within the chamber and measures the temperature and humidity of the
air as it passes over the measuring head. As with climatic chambers, it is assumed that, at equilibrium,
water vapour pressure is homogenous throughout the chamber volume. Provided that the temperature
is stable and uniform, the %RH derived from the dew point and temperature measurement is suitable
for use as a transfer standard for calibration of the generator
’s control probe or as a reference against
which units under test can be compared.
Temperature Effects
Most RH generators are also able to control temperature. This
means that RH probes can be calibrated for temperature at the
same time as RH, and RH calibration can be performed at
working temperature. As RH is highly temperature dependent,
temperature effects and gradients in particular, must be
considered. With the RP2 probe installed into the RH generator
chamber, the Peltier element that is used to cool the mirror
causes heat to be dissipated from the body of the measuring
head. This can cause temperature gradients within the generator
chamber, so these effects should be evaluated and
consequential uncertainty contributions applied to an uncertainty
budget.
Summary of Contents for 473-RP2
Page 1: ...Operation and Maintenance Manual V5 0 Dew Point Mirror 473...
Page 2: ......
Page 4: ...ii MBW 473 Manual English V5 0...
Page 8: ...2 MBW 473 Manual English V5 0...
Page 10: ...4 MBW 473 Manual English V5 0...
Page 14: ...8 MBW 473 Manual English V5 0 SH2 SHX Fan Fan...
Page 30: ...24 MBW 473 Manual English V5 0...
Page 68: ...62 MBW 473 Manual English V5 0...
Page 69: ...MBW 473 Manual English V5 0 63 11 Drawings 11 1 Display Unit Side View...
Page 70: ...64 MBW 473 Manual English V5 0 Front View Back View...
Page 71: ...MBW 473 Manual English V5 0 65 11 2 Measuring Heads RP2 SH2...
Page 72: ...66 MBW 473 Manual English V5 0 SHX SH3...
Page 73: ...MBW 473 Manual English V5 0 67...
Page 74: ...68 MBW 473 Manual English V5 0...