MBW 473 Manual English V5.0
Analog Outputs
If the analog output option is fitted to the 473, it can be configured in this menu.
Diagnostic Functions
Here you have access to the System Info, Ice Test, Peltier Cooling Test (PCT) and baud
rate settings for the RS232 port.
5.2 Selection of indicated Parameters
In the
Parameter
menu you can choose which parameters you would like to have displayed on the
data lines. When you select parameters for display on any of the four data lines, those selections re-
main valid until you change them again, even after you turn the 473 off. Below you will find the list of
the available parameters.
Parameter
Explanation
Dew Point
The temperature to which a gas must be cooled to start condensing water vapor to liquid water.
Dew point is pressure dependent and must be stated together with its associated pressure.
Frost Point
The temperature to which a gas must be cooled to start deposition of water vapor in the
form of ice. Frost point is pressure dependent and must be stated together with its associ-
ated pressure. Frost point exists only below 0 °C. While not technically correct, it has been
common industry practice to report values below 0 °C as dew point, although frost point is
the correct term. For further explanation on
dew or frost point refer to ‘Dew / Frost Control’
in section 5.4.
%RH
The ratio between the amount of water vapor in a sample and the maximum amount possi-
ble at that same temperature and pressure.
%RH WMO
The ratio between the amount of water vapor in a sample and the maximum amount possi-
ble at that same temperature and pressure calculated using the World Meteorological Or-
ganization (WMO) formula.
Volume Ratio
The ratio between the water vapor volume and the total volume of the sample gas, general-
ly expressed in parts per million by volume, ppm
v
or its numerical
equivalent μl/l. Once
determined, ppm
v
has no further pressure dependence. It is also independent of the gas
type or mixture.
Weight Ratio
Weight ratio is the ratio between the mass of water vapor and the total mass of the sample
gas, generally expressed in parts per million by weight, ppm
w
or its numerical equivalent
mg/kg. Once determined, ppm
w
has no further pressure dependence, but depends on the
gas type and mixture through the molecular weight of the constituents.
Absolute Humidity
The weight of water vapor per unit volume of humidified gas.
Specific Humidity
A ratio of the water vapor to the total weight of the humidified gas.
Vapor Pressure
The partial pressure exerted by vapor in thermodynamic equilibrium with its condensed
phases (solid or liquid) at a given temperature. It is usually expressed in kPa.
Head Pressure
The pressure of the gas sample in the measuring head.
External Temp
The temperature measured by the external temperature probe.
Head Temp
The temperature measured by the PRT in the measuring head.
Summary of Contents for 473-RP2
Page 1: ...Operation and Maintenance Manual V5 0 Dew Point Mirror 473...
Page 2: ......
Page 4: ...ii MBW 473 Manual English V5 0...
Page 8: ...2 MBW 473 Manual English V5 0...
Page 10: ...4 MBW 473 Manual English V5 0...
Page 14: ...8 MBW 473 Manual English V5 0 SH2 SHX Fan Fan...
Page 30: ...24 MBW 473 Manual English V5 0...
Page 68: ...62 MBW 473 Manual English V5 0...
Page 69: ...MBW 473 Manual English V5 0 63 11 Drawings 11 1 Display Unit Side View...
Page 70: ...64 MBW 473 Manual English V5 0 Front View Back View...
Page 71: ...MBW 473 Manual English V5 0 65 11 2 Measuring Heads RP2 SH2...
Page 72: ...66 MBW 473 Manual English V5 0 SHX SH3...
Page 73: ...MBW 473 Manual English V5 0 67...
Page 74: ...68 MBW 473 Manual English V5 0...