30
MBW 473 Manual English V5.0
5.4 Control Setup
Using the
Control Setup
menu enables you to control the manner in which the 473 operates.
Dew / Frost Control
When measuring dew/frost points between 0 °C and -
20 °C, condensation on the instrument’s chilled
mirror may be in the form of dew, frost, or a combination of both. If the state of the condensation is
not known, it will introduce errors into all the humidity measurements reported by the instrument.
To eliminate this potential source of error, the Force Frost function is used to rapidly cool the mirror to
below -20 °C, forcing all dew on the mirror into frost. The mirror will then re-stabilize at the frost point
temperature. Once the condensate layer is in a state of frost, it will remain frost for all sub-zero mirror
temperatures, allowing the instrument to measure the frost point accurately. The dew point and all
other humidity measurements are then mathematically calculated from the frost point.
To change the
Force Frost
settings:
1. Select the
Control Setup
menu by
pressing the dark gray menu selection
key until
Control Setup
appears.
2. Touch the
Dew/Frost Control
button.
The Mirror
Dew/Frost Control
window
will open.
3. The Force Frost function can be enabled or disabled by clicking on the
Force Frost Below
button.
If the indicator on the left side of the button is green, Force Frost is enabled. If the indicator is
grey, Force Frost is disabled.
4. To adjust the temperature below which Force Frost activates, click on the
Change To
: field to the
right of the Force Frost Below button. The field will turn white.
5. Enter the temperature in degrees C below which Force Frost should activate.
6. If you make an entry error, touch the field that holds the number you wish to change. Each time
you touch the field, the last digit entered will be deleted.
Press the
OK
button in the dialog box or press Enter on the keypad to confirm the new value. Press
Cancel to leave it unchanged.
Summary of Contents for 473-RP2
Page 1: ...Operation and Maintenance Manual V5 0 Dew Point Mirror 473...
Page 2: ......
Page 4: ...ii MBW 473 Manual English V5 0...
Page 8: ...2 MBW 473 Manual English V5 0...
Page 10: ...4 MBW 473 Manual English V5 0...
Page 14: ...8 MBW 473 Manual English V5 0 SH2 SHX Fan Fan...
Page 30: ...24 MBW 473 Manual English V5 0...
Page 68: ...62 MBW 473 Manual English V5 0...
Page 69: ...MBW 473 Manual English V5 0 63 11 Drawings 11 1 Display Unit Side View...
Page 70: ...64 MBW 473 Manual English V5 0 Front View Back View...
Page 71: ...MBW 473 Manual English V5 0 65 11 2 Measuring Heads RP2 SH2...
Page 72: ...66 MBW 473 Manual English V5 0 SHX SH3...
Page 73: ...MBW 473 Manual English V5 0 67...
Page 74: ...68 MBW 473 Manual English V5 0...