22
MBW 473 Manual English V5.0
The RP2 Measuring Head
1. Temperature Probe Ø3 x 30 mm
2. Measuring Head Assembly
3. 19-pin Measuring Head Cable (2 m)
4. When reassembling the measuring head
take care to align the red marks.
5. External Temperature Connection
6. Polished Mirror
The 473 instrument series has three different display units depending on the type of re-
mote cable mounted measuring head:
- 473-SH2 /-RP2
- 473-SH3
- 473-SHX
Use only the corresponding measuring head for each display unit, otherwise it will cause damage to
the instrument. The SHX version is equipped with a 30-pin measuring head connection instead of the
19-pin connection used on the other versions. This makes it impossible to accidentally fit the SHX
measuring head to the other display units and vice versa. However, the SH3, SH2 and RP2 use the
same 19-pin connection. Therefore, caution must be taken not to connect the SH3 measuring head to
the 473-SH2 /-RP2 display unit or the SH2 or RP2 measuring heads to the 473-SH3 display unit.
1
2
3
4
5
6
Summary of Contents for 473-RP2
Page 1: ...Operation and Maintenance Manual V5 0 Dew Point Mirror 473...
Page 2: ......
Page 4: ...ii MBW 473 Manual English V5 0...
Page 8: ...2 MBW 473 Manual English V5 0...
Page 10: ...4 MBW 473 Manual English V5 0...
Page 14: ...8 MBW 473 Manual English V5 0 SH2 SHX Fan Fan...
Page 30: ...24 MBW 473 Manual English V5 0...
Page 68: ...62 MBW 473 Manual English V5 0...
Page 69: ...MBW 473 Manual English V5 0 63 11 Drawings 11 1 Display Unit Side View...
Page 70: ...64 MBW 473 Manual English V5 0 Front View Back View...
Page 71: ...MBW 473 Manual English V5 0 65 11 2 Measuring Heads RP2 SH2...
Page 72: ...66 MBW 473 Manual English V5 0 SHX SH3...
Page 73: ...MBW 473 Manual English V5 0 67...
Page 74: ...68 MBW 473 Manual English V5 0...