Automatic Mirror Check
To view or edit the Mirror Check parameters, press the
Mirror Check
key of the
Control Setup
menu.
If automatic mirror checks are desired, se-
lect it by pressing the
Cycle Time
button. A
green light on the button indicates that au-
tomatic mirror check is enabled.
When the automatic mirror check is enabled, the
Mirror Check
key at the bottom
of the screen shows a countdown timer indicating the time before the next auto-
matic mirror check is performed. In the automatic mode, the mirror check may
still be initiated manually by pressing the
Mirror Check
button.
Cycle Time
The
Cycle Time
is the number of minutes between automatic mirror check operations. Use the numer-
ical keypad to enter the desired cycle time in minutes.
Heating Time
The Heating Time determines how long the mirror check temperature will be held before allowing the
next dew or frost point measurement. A heating time of 0 means that the instrument will resume dew
or frost point measurement immediately upon reaching the mirror check temperature. If a heating time
greater than 0 is entered, the mirror will heat and remain at that temperature for the chosen duration.
Heating time is effective regardless of whether mirror check is triggered automatically or manually.
Temperature
Edit the
Temperature
field to change the temperature, in degrees Celsius, that the mirror will be heat-
ed to, and optionally held at during Mirror Check.
If you have entered a wrong value into a field and want to erase it, press the entry field to
backspace.
Fan Settings
The 473-SH2 and 473-SHX versions include a fan that allows the instrument to extract a sample from
the application for measurement. The fan has variable power
(range 20…100%) so that the user can
change the flow rate. It is recommended that the fan
power is set within the range 20…40% for typical
use, that is with no or little tubing in front of the measuring head.
Summary of Contents for 473-RP2
Page 1: ...Operation and Maintenance Manual V5 0 Dew Point Mirror 473...
Page 2: ......
Page 4: ...ii MBW 473 Manual English V5 0...
Page 8: ...2 MBW 473 Manual English V5 0...
Page 10: ...4 MBW 473 Manual English V5 0...
Page 14: ...8 MBW 473 Manual English V5 0 SH2 SHX Fan Fan...
Page 30: ...24 MBW 473 Manual English V5 0...
Page 68: ...62 MBW 473 Manual English V5 0...
Page 69: ...MBW 473 Manual English V5 0 63 11 Drawings 11 1 Display Unit Side View...
Page 70: ...64 MBW 473 Manual English V5 0 Front View Back View...
Page 71: ...MBW 473 Manual English V5 0 65 11 2 Measuring Heads RP2 SH2...
Page 72: ...66 MBW 473 Manual English V5 0 SHX SH3...
Page 73: ...MBW 473 Manual English V5 0 67...
Page 74: ...68 MBW 473 Manual English V5 0...