MBW 473 Manual English V5.0
21
This picture shows a possible set up
for the SH3 measuring head. A stain-
less steel tube is connected to the
gas inlet. At the gas outlet there is a
mechanical flow meter installed to
regulate and measure the gas flow
rate. We recommend 0.5 - 1 lpm for
frost point measurements. A flow
rate of 0.5 lpm is suitable for dew
point measurements.
Please note that connecting tubes as
well as flow meter are not included
with the instrument.
SH3 Fan
One of the differences of the SH3 measuring head to the SH2 measuring head is that it does not have
a sample gas flow fan. The purpose of the fan installed on the SH3 is to cool the measuring head
base. The fan dissipates the heat generated by the Peltier during operation. This is especially im-
portant when measuring low frost points. To prevent the measuring head from overheating the fan
runs continuously during operation and cannot be switched off. The
Fan
button on the display refers to
the gas flow fan of the SH2. Therefore, when you have a SH3 measuring head installed and you press
the
Fan
button, a message will appear saying you that there is no fan installed.
Summary of Contents for 473-RP2
Page 1: ...Operation and Maintenance Manual V5 0 Dew Point Mirror 473...
Page 2: ......
Page 4: ...ii MBW 473 Manual English V5 0...
Page 8: ...2 MBW 473 Manual English V5 0...
Page 10: ...4 MBW 473 Manual English V5 0...
Page 14: ...8 MBW 473 Manual English V5 0 SH2 SHX Fan Fan...
Page 30: ...24 MBW 473 Manual English V5 0...
Page 68: ...62 MBW 473 Manual English V5 0...
Page 69: ...MBW 473 Manual English V5 0 63 11 Drawings 11 1 Display Unit Side View...
Page 70: ...64 MBW 473 Manual English V5 0 Front View Back View...
Page 71: ...MBW 473 Manual English V5 0 65 11 2 Measuring Heads RP2 SH2...
Page 72: ...66 MBW 473 Manual English V5 0 SHX SH3...
Page 73: ...MBW 473 Manual English V5 0 67...
Page 74: ...68 MBW 473 Manual English V5 0...