Entrée Bath Gen III -
MODEL MB-68 AV-XL
MAINTENANCE MANUAL
10-20-19
Page 39 of 51
450 E 32
nd
Ave – Fremont, NE 68025
800-798-5867 FAX: 402-721-3023
www.mastercarebath.com
MasterCare hygiene products are designed and manufactured to help provide a more operationally efficient and secure
environment for your staff and residents. It is required that a scheduled maintenance program be set up to insure that
the efficiencies and security designed into the bathing system are kept at peak levels. Instructions on how to set up a
scheduled maintenance program are covered in this manual.
Maintenance Instructions:
MasterCare Bathing Systems are purposely designed to minimize the maintenance requirements associated with such
products. They are manufactured using materials suited to their application and the durability requirements of the health
care industry.
Daily:
•
Check for loose fasteners, worn seat belts and pads, leaks on the floor and mixing valve temperature.
Weekly:
•
Test for proper operation of the GFCI.
Monthly:
•
Door assembly: Check all components on the entire door assembly to ensure they are properly secured and not
missing. Check the security belt and pads for signs of excessive wear. Correct all deficiencies found during the
inspections before putting unit back into service.
Quarterly:
•
Door assembly: Same as monthly.
•
Plumbing and drain assembly: Check all fittings and connections to make sure they are secure and no leaks have
developed. Check with a hand held thermometer and observe the solar thermometer of incoming water to
ensure it is accurate.
•
BathAire
(SM)
system: Check all air supply pipes and hoses in the system so they are secure and no air leaks have
developed. Also check to make certain no water leaks have developed.
•
Control system: Check all controls to ensure that they function as outlined in the operations manual.
CAUTION
: All deficiencies discovered during monthly and quarterly inspections must be corrected and/or
repaired immediately. This will help prevent potential damage to the system that could occur if corrective
action is not taken and ultimately extend the life of the bathing system.
CAUTION:
The caution heading identifies important
maintenance and operation information. Failure to
obey a caution warning may result in damage to the
Entree Bath and may void the warranty.
Before performing any checks that require
inspection inside of the tub shell, trip the
GFCI circuit breaker to OFF. It is located at
the Head-End of the tub
.
IMPORTANT SAFETY INSTRUCTIONS.
READ AND FOLLOW ALL INSTRUCTIONS
THROUGHLY
Failures of materials can occur due to normal
wear associated with use or misuse over time.
It is in the best interests of the residents, staff
and the provider that recommended safety
inspections are scheduled on a periodic basis
and preventive maintenance performed as
needed.