VR2250
Page 22 of 131
6.2.2. Results of "MAN" mode
At the end of the test if
the Insulation resistance
is
lower than R
MAX
(see
Table2
)
and the
instrument generated
the Nominal test
Voltage
, the instrument
emits a
double sound
signal
indicating the
positive outcome of
the test
and displays
one screen similar to the
screen alongside.
M
Ω
05.06.01
Insulation Resistance
1.07
M
Ω
514V 15s
MAN 500V
Voltage during the Test
Duration of the Test
Test mode
Test voltage set
FUNC
VNOM
In order to evaluate the test you must compare the result with the limits indicated in the
Guidelines (see Table1).
The displayed result can be stored pressing the
SAVE
key
twice
(according to
paragraph 9.1).
If the Insulation
resistance is
higher
than R
MAX
(see
Table2
),
the instrument emits a
double sound signal
at
the end of the test
indicating the positive
outcome of the test
and displays one screen
similar to the screen
alongside.
M
Ω
05.06.01
Maximum resistance value
which can be measured
(999
Ω
is displayed if a
rated voltage of 500V was
selected see Table2).
> 999
M
Ω
523V 15s
MAN 500V
The symbol
">"
means that
the resistance value is
higher than R
MAX
.
Test
duration
FUNC
VNOM
The displayed result can be stored pressing the
SAVE
key
twice
(according to
paragraph 9.1).