6.1.4. Results of "RT+" and "RT-" modes
if a
resistance value
RT+
or
RT+ lower than
5
Ω
is detected, the
instrument emits a
double sound signal
indicating the positive
outcome
of the test and
displays one screen
similar to the screen
alongside.
LOW
Ω
05.06.01
Max Resistance value of
R+
or
R-
.
1.07
Ω
219mA
RT+ 0.11
Ω
TIME: 10s
Test current
Duration of the Test
FUNC
CAL
↑
↓
Note
:
We recommend that crocodile clips are used to attain a good contact with the
conductor under test. Indeed in this test the instrument gives as final result the
maximum measured value of R+ or R- and using test leads instead of crocodile clips
could give you a faulty result due to a faulty contact between the test leads and
conductor under test.
The displayed result can be stored pressing the
SAVE
key
twice
(according to
paragraph 9.1).
6.1.5. "AUTO", RT+", "RT-" faulty cases
If the instrument detects
the External Power
supply adapter
connected to the
instrument will display
the screen along side.
LOW
Ω
05.06.01
-.- -
Ω
R+ R-
---
Ω
---
Ω
---mA ---mA
REMOVE POWER
AUTO 0.11
Ω
Disconnect the External
Power Supply Adapter
FUNC CAL
If the terminal voltage is
higher than 15V, the
instrument does not
carry out the test and
displays the screen
alongside for 5 seconds.
LOW
Ω
05.06.01
-.- -
Ω
R+ R-
-.--
Ω
-.--
Ω
---mA ---mA
VOLT IN INPUT
AUTO 0.11
Ω
ATTENTION:
the test was
not performed
because of
voltage at the terminal
ends.
FUNC