VR2250
Page 70 of 131
Symbols Description
Advised
settings
START:MAN
The recording of all the selected parameters will start at 00
seconds after pressing
START/STOP
(see chapter REF
_Ref6712586 \r \h 10.1).
☺
STOP:MAN
The recording of all the selected parameters will be interrupted
manually by pressing
START/STOP
(see chapter 10.1).
☺
START:AUTO
STOP:AUTO
The recording of all the selected values will be started / interrupted at
the set dates and times. In order to start the recording the user will
have to press START/STOP to set the instrument in Stand-by mode
until the start date and time previously set (see chapter REF
_Ref6712622 \r \h 10.1).
INT. PERIOD
The value of this parameter determines how many seconds will
elapse before the values of
all the selected parameters
will be
memorised (see chapter 16.13.1). Available choices:
5sec,10sec,30sec,1min,2min,5min,10min,15min,30min,60min.
15min
☺
HARM REC.
ON
= the instrument will record the values of the selected
voltage and current harmonics
.
☺
OFF = the instrument
will not record
any voltage or current
harmonic selected
ANOM REC.
ON
= the Instrument will record Voltage Anomalies (voltage Sag
and Surge) (see paragraph 16.10)
☺
OFF = the instrument
will not record
any voltage Sag and Surge
V1 Voltage
RMS
value
☺
V1
THD, DC, 01...49
Voltage Total Harmonic Distortion, DC Component, 01..49
Harmonics respectively.
☺
THD,01,03,05,07
Vref
(Only if ANOM. REC
flag has been set ON)
RMS reference value for Voltage used in Voltage Anomalies
detection (Voltage Sag and Surge). The Reference is:
a) Voltage Phase to Neutral for Single Phase and 4 wire three
phase system.
b) Voltage Phase to Phase for 3 wire three phase system.
☺
Single phase:
230V
3 Phase system:
3 wires:
400V
4 wires
230V
LIM+, LIM-
(Only if ANOM. REC
flag has been set ON)
High and Low Voltage Percent threshold used in Voltage
Anomalies detection (Voltage Sag and Surge). These parameters
can be adjusted in range 3%
÷
30% (step 1%).
Example: Three Phase System 4 wire.
Vref = 230, LIM+= 6%, LIM-=10% =>
High Lim = 243.8V, Low Lim = 207.0V
The Instrument will detect a voltage Anomaly if the RMS Voltage
Values (calculated every 10ms) are beyond the above calculated
thresholds (see paragraph 16.10.
☺
+6% / -10%
I1
Current RMS value.
☺
I1
THD, DC, 01..49
Current Total Harmonic Distortion, DC Component, 01..49
Harmonics respectively
☺
THD,01,03,05,07