Standard condition
specification
Inaccuracy in this document is given under the following
standard conditions and for the worst case, i.e. all errors
are summarized positive. Statistically 95% of all values
will be within 2/3 of the given inaccuracy.
• Ambient pressure: 101.3 kPa
• Room temperature: 20 ºC
• Dry gases in patient system
• Inlet pressure: 4.3 kPa x 100
• Pre-use check performed on a warmed up ventilator
• Default settings unless otherwise specified
Inspiratory channel
Pressure drop Max. 6 cmH
2
O at a flow of 1 l/s
Internal compressible factor Max. 0.1 ml/cmH
2
O
Gas delivery system Microprocessor controlled valves
Gas delivery device
Flow range:
Adult
0-3.3 l/s
Infant
0-0.55 l/s
Inaccuracy:
±5% or ± 0.1 ml/s
Pressure setting:
Max. 80/120 cmH
2
O (Infant/Adult)
Inaccuracy:
± 5% or ± 1 cmH
2
O
1
NIV Max leakage compensation level:
Adult
50 l/min
Infant
15 l/min
Infant Nasal CPAP 10 l/min
O
2
concentration
Setting range:
21 - 100%
Inaccuracy:
± 3% O
2
Inspiratory Minute Volume
Adult
Setting range:
0.5 - 60 l/min
Inaccuracy:
± 6%
2
Infant
Setting range:
0.3 - 20 l/min
Inaccuracy:
± 6%
3
Inspiratory Tidal Volume
Adult
Setting range:
100 - 2000/4000 ml
Inaccuracy:
± 7%
4
Infant
Setting range:
2 - 350 ml
5
Inaccuracy: ±
6%
6
Expiratory channel
Pressure drop Max. 3 cmH
2
O at a flow of 1 l/s
Internal compressible factor Max. 0.1 ml/cmH
2
O
PEEP regulation Microprocessor controlled valve
PEEP setting range:0 - 50 cmH
2
O
Inaccuracy:
± 5% or ± 1 cmH
2
O
7
Expiratory flow measurements
Flow range:
0 - 3.2 l/s
Rise time:
< 12 ms for 10 - 90% response at a
flow of 0.05 - 3.2 l/s.
Inaccuracy:
± 5% or ± 2.5 ml/s
Monitoring
Expiratory Minute Volume
Adult
Range:
0 - 60 l/min
Inaccuracy:
± 8% or ± 0.15 l/min
8
NIV
± 10%
9
Infant
Range:
0 - 20 l/min
Inaccuracy:
± 8% or ± 0.15 l/min
10
NIV
± 10%
11
NIV, Nasal CPAP
± 25% or ± 0.15 l/min
12
Expiratory Tidal Volume
Adult
Range:
0 - 2000/4000 ml
Inaccuracy:
± 8% or ± 18 ml
13
Infant
Range:
0 - 350 ml
Inaccuracy:
± 8% or ± 2 ml
14
O
2
concentration
Range:
0 - 100%
Inaccuracy:
± 5% of read value
Airway pressure
Range:
-40 - 160 cmH
2
O
Inaccuracy:
± 5% or ± 1 cmH
2
O
Supply pressure
Range:
0 - 7 bar
Inaccuracy:
± 5% of read value
1.
at RR < 100 b/min
2.
at 2.5-60 l/min and set I:E<1:1
3.
at 1-20 l/min and set I:E<1:1
4.
at 400-4000 ml and set I:E<1:1
5.
Infant: tidal volume range in volume
controlled/supported modes is 2-350 ml.
When the tidal volume is set below 5 ml,
use of Y Sensor measuring is
recommended to optimize monitoring of
volumes.
6.
at 20-350 ml and set I:E<1:1
7.
at RR < 60 b/min
8.
at RR < 100 b/min
9.
at constant leakage fraction <30%
10. at RR < 100 b/min
11. at constant leakage fraction <30%
12. at constant leakage fraction <30%
13. at Expiration time < 4 s and RR <100 b/min
14. at Expiration time < 1 s and RR <100 b/min
Technical data
Summary of Contents for SERVO-i V3.1
Page 1: ...User s manual US Version VENTILATOR SYSTEM SERVO i V3 1 ...
Page 2: ......
Page 4: ...2 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options ...
Page 16: ...14 Servo User s manual US edition Order No 66 00 261 1 Infant Adult Universal Options Notes ...
Page 71: ...69 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 2 Notes ...
Page 72: ...70 Servo User s manual US edition Order No 66 00 261 2 Infant Adult Universal Options Notes ...
Page 81: ...79 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 3 Notes ...
Page 82: ...80 Servo User s manual US edition Order No 66 00 261 3 Infant Adult Universal Options Notes ...
Page 120: ...118 Servo User s manual US edition Order No 66 00 261 4 Infant Adult Universal Options Notes ...
Page 146: ...144 Servo User s manual US edition Order No 66 00 261 5 Infant Adult Universal Options Notes ...
Page 157: ...155 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 6 Notes ...
Page 158: ...156 Servo User s manual US edition Order No 66 00 261 6 Infant Adult Universal Options Notes ...
Page 191: ...189 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 7 Notes ...
Page 192: ...190 Servo User s manual US edition Order No 66 00 261 7 Infant Adult Universal Options Notes ...
Page 211: ...209 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 8 Notes ...
Page 212: ...210 Servo User s manual US edition Order No 66 00 261 8 Infant Adult Universal Options Notes ...
Page 225: ...223 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 9 Notes ...
Page 226: ...224 Servo User s manual US edition Order No 66 00 261 9 Infant Adult Universal Options Notes ...
Page 242: ...240 Servo User s manual US edition Order No 66 00 261 10 Infant Adult Universal Options Notes ...
Page 256: ...254 Servo User s manual US edition Order No 66 00 261 11 Infant Adult Universal Options Notes ...
Page 259: ...257 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 12 Notes ...
Page 260: ...258 Servo User s manual US edition Order No 66 00 261 12 Infant Adult Universal Options Notes ...
Page 274: ...272 Servo User s manual US edition Order No 66 00 261 13 Infant Adult Universal Options Notes ...
Page 278: ...276 Servo User s manual US edition Order No 66 00 261 14 Infant Adult Universal Options Notes ...
Page 279: ......
Page 283: ...VENTILATOR SYSTEM V 3 1 USER S MANUAL US VERSION SERVO i ...