13
Servo
…
User´s manual
US edition
Order No: 66 00 261
Infant
Adult
Universal
Options
1
Nebulization
• Servo Humidifier/HME must be
disconnected during nebulization
otherwise the humidifier may be blocked
(page 128).
• The heated humidifier must be switched
off during nebulization. Otherwise the
particle size may be affected (page 128).
• During nebulization a filter must be
connected to the expiratory inlet of the
ventilator. Always carefully monitor the
airway pressure during nebulization.
Increased airway pressure could be
caused by a clogged filter. The filter should
be replaced if the expiratory resistance
increases or every 24 hours when the
nebulizer is being used.
• When a Servo Ultra Nebulizer is used,
always consult the drug manufacturer
regarding the appropriateness of
ultrasonic nebulization for certain
medications (page 128, 187).
• The nebulizer must not be used without
buffer liquid (sterile water). Otherwise the
ultrasonic generator crystal may break
(page 129, 187).
• To avoid explosion hazards, flammable
agents such as ether and cyclopropane
must not be used with this device. Only
agents which comply with the
requirements on non-flammable agents in
the IEC standard “Particular requirements
for electrical safety of anaesthetic
machines” are suitable.
• For adult/pediatric patients, never fill the
medication cup with more than 10 ml
(page 129).
• For neonatal patients, never fill the
medication cup with more than 4 ml
(page 129).
• If the patient unit of the nebulizer is tilted,
the drug can flow into the patient´s lungs
or the ventilator.
• The nebulizer must not be left unattended
when connected to a patient.
• Continuously check that the buffer liquid
level is between MIN. and MAX. during
nebulization (page 187).
• During nebulization: Continuously check
that moisture is generated in the
medication cup (page 187).
• When the ventilator is running on batteries
the nebulizer module is inoperative, to
reduce the power consumption
(page 187).
• For information about the stand alone
Aeroneb Professional Nebulizer System,
refer to separate manual.
Cleaning
• All personnel should be aware of the risk
of parts being infected when
disassembling and cleaning the ventilator
(page 191).
• After removing the Expiratory cassette, do
not pour any fluid into the Expiratory
cassette compartment (page 196).
Replacement of O
2
cell
The sealed unit of the O2 cell, contains a
caustic liquid which may cause severe burns
to the skin and eyes. In case of contact,
immediately flush continuously with water for
at least 15 minutes and seek medical
attention especially if the eyes are affected
(page 214)
Warnings, cautions and important
Summary of Contents for SERVO-i V3.1
Page 1: ...User s manual US Version VENTILATOR SYSTEM SERVO i V3 1 ...
Page 2: ......
Page 4: ...2 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options ...
Page 16: ...14 Servo User s manual US edition Order No 66 00 261 1 Infant Adult Universal Options Notes ...
Page 71: ...69 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 2 Notes ...
Page 72: ...70 Servo User s manual US edition Order No 66 00 261 2 Infant Adult Universal Options Notes ...
Page 81: ...79 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 3 Notes ...
Page 82: ...80 Servo User s manual US edition Order No 66 00 261 3 Infant Adult Universal Options Notes ...
Page 120: ...118 Servo User s manual US edition Order No 66 00 261 4 Infant Adult Universal Options Notes ...
Page 146: ...144 Servo User s manual US edition Order No 66 00 261 5 Infant Adult Universal Options Notes ...
Page 157: ...155 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 6 Notes ...
Page 158: ...156 Servo User s manual US edition Order No 66 00 261 6 Infant Adult Universal Options Notes ...
Page 191: ...189 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 7 Notes ...
Page 192: ...190 Servo User s manual US edition Order No 66 00 261 7 Infant Adult Universal Options Notes ...
Page 211: ...209 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 8 Notes ...
Page 212: ...210 Servo User s manual US edition Order No 66 00 261 8 Infant Adult Universal Options Notes ...
Page 225: ...223 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 9 Notes ...
Page 226: ...224 Servo User s manual US edition Order No 66 00 261 9 Infant Adult Universal Options Notes ...
Page 242: ...240 Servo User s manual US edition Order No 66 00 261 10 Infant Adult Universal Options Notes ...
Page 256: ...254 Servo User s manual US edition Order No 66 00 261 11 Infant Adult Universal Options Notes ...
Page 259: ...257 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 12 Notes ...
Page 260: ...258 Servo User s manual US edition Order No 66 00 261 12 Infant Adult Universal Options Notes ...
Page 274: ...272 Servo User s manual US edition Order No 66 00 261 13 Infant Adult Universal Options Notes ...
Page 278: ...276 Servo User s manual US edition Order No 66 00 261 14 Infant Adult Universal Options Notes ...
Page 279: ......
Page 283: ...VENTILATOR SYSTEM V 3 1 USER S MANUAL US VERSION SERVO i ...