228
Servo
…
User´s manual
US edition
Order No: 66 00 261
10
Infant
Adult
Universal
Options
O
2
concentration: Low
Measured O
2
concentration is below
the set value by more than 6 Vol.%,
or concentration below 18 Vol.%
which is independent of operator
settings. Refer to page 177 for O
2
cell adaptation.
Gas delivered in O
2
supply line is not
O
2
. O
2
sensor faulty or exhausted.
O
2
cell uncalibrated. O
2
/oxygen gas
module faulty.
Check O
2
supply line. Perform a
Pre-use check.
Refer to page 177 for O
2
cell
adaptation.
Paw high
Caution: If airway
pressure rises 6 cmH
2
O
above set upper
pressure limit, the safety
valve opens.The safety
valve also opens if
system pressure
exceeds 117± 7 cmH
2
O.
Airway pressure exceeds preset
Upper pressure limit.
Kinked or blocked tubing. Mucus or
secretion plug in endotracheal tube
or in airways. Patient coughing or
fighting ventilator. Inspiratory flow
rate too high, Improper alarm
setting. Blocked expiratory filter.
Check patient and breathing
system.
Check ventilator settings and alarm
limits.
Restart ventilator!
Software related error.
Restart the ventilator and perform a
Pre-use check. If the problem still
remains, take the unit out of
operation and contact a service
technician.
Safety valve test failed
During Pre-use check the system
found failures during the check of
the opening pressure for the safety
valve.
Contact a service technician.
Settings lost; Restart
ventilator
Software error, memory corrupt.
Restart the ventilator and perform a
Pre-use check.
Check ventilator settings.
Technical error in
Expiratory cassette
Technical problem with the
Expiratory cassette.
Perform a Pre-use check. Change
the Expiratory cassette and perform
a Pre-use check. If the problem still
remains contact a service
technician.
Technical error: Restart
ventilator
Ventilator settings lost.
Restart the ventilator, perform a
Pre-use check and check all
settings. If the problem still remains
contact a service technician.
Time in waiting position
exceeds 2 min.
Time in waiting position is exceeded.
Patient is not connected to the
ventilator or leakage is excessive.
Check patient and breathing
system.
Problem
(Display message)
Possible cause
Remedy
High priority alarms
Summary of Contents for SERVO-i V3.1
Page 1: ...User s manual US Version VENTILATOR SYSTEM SERVO i V3 1 ...
Page 2: ......
Page 4: ...2 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options ...
Page 16: ...14 Servo User s manual US edition Order No 66 00 261 1 Infant Adult Universal Options Notes ...
Page 71: ...69 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 2 Notes ...
Page 72: ...70 Servo User s manual US edition Order No 66 00 261 2 Infant Adult Universal Options Notes ...
Page 81: ...79 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 3 Notes ...
Page 82: ...80 Servo User s manual US edition Order No 66 00 261 3 Infant Adult Universal Options Notes ...
Page 120: ...118 Servo User s manual US edition Order No 66 00 261 4 Infant Adult Universal Options Notes ...
Page 146: ...144 Servo User s manual US edition Order No 66 00 261 5 Infant Adult Universal Options Notes ...
Page 157: ...155 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 6 Notes ...
Page 158: ...156 Servo User s manual US edition Order No 66 00 261 6 Infant Adult Universal Options Notes ...
Page 191: ...189 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 7 Notes ...
Page 192: ...190 Servo User s manual US edition Order No 66 00 261 7 Infant Adult Universal Options Notes ...
Page 211: ...209 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 8 Notes ...
Page 212: ...210 Servo User s manual US edition Order No 66 00 261 8 Infant Adult Universal Options Notes ...
Page 225: ...223 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 9 Notes ...
Page 226: ...224 Servo User s manual US edition Order No 66 00 261 9 Infant Adult Universal Options Notes ...
Page 242: ...240 Servo User s manual US edition Order No 66 00 261 10 Infant Adult Universal Options Notes ...
Page 256: ...254 Servo User s manual US edition Order No 66 00 261 11 Infant Adult Universal Options Notes ...
Page 259: ...257 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 12 Notes ...
Page 260: ...258 Servo User s manual US edition Order No 66 00 261 12 Infant Adult Universal Options Notes ...
Page 274: ...272 Servo User s manual US edition Order No 66 00 261 13 Infant Adult Universal Options Notes ...
Page 278: ...276 Servo User s manual US edition Order No 66 00 261 14 Infant Adult Universal Options Notes ...
Page 279: ......
Page 283: ...VENTILATOR SYSTEM V 3 1 USER S MANUAL US VERSION SERVO i ...