Infant
Adult
Universal
Options
10
CO
2
sensor error
Hardware error in CO
2
Capnostat
sensor. The values in the Capnostat
memory failed the internal test.
Check that the Capnostat sensor is
properly plugged in. Re insert the
Capnostat sensor if necessary.
Calibrate the Capnostat sensor
(refer to page 151). Change the
Capnostat sensor. If the error
persists, contact a service
technician.
CO
2
sensor temperature
too high
Possible hardware error. The
Capnostat sensor temperature is
higher than 50
o
C.
Make sure that the Capnostat
sensor is not exposed to extreme
heat (heat lamp, incubator etc.). If
the error persist, the problem is
most likely a faulty Capnostat
sensor.
CO
2
sensor temperature
too low
The Capnostat sensor does not
reach operating temperature.
The problem is probably hardware
related. Change Capnostat sensor
and/or module. Call a service
technician.
CPAP High/Low
Preset or default alarm exceeded.
Check patient and breathing
system. Check mas k/ prongs size
and accurate adaptation to the
patient
Check alarm settings.
etCO
2
high
Hypoventilation. Leakage with high
bias flow. CO
2
sensor, Y-piece,
HME
Check patient circuit. Ventilator
settings.
etCO
2
low
Hyperventilation. Leakage with high
bias flow. CO
2
sensor, Y-piece,
HME
Check patient circuit. Ventilator
settings.
Exp. cassette exchanged Expiratory cassette has been
exchanged during operation. No
Pre-use check performed after
exchange.
Perform a Pre-use check.
Inspiratory flow
overrange
Combination of settings exceeds the
allowable inspiration flow range.
Change the ventilator settings.
Increase the gas inlet pressure.
Internal temperature:
High
Temperature inside the ventilator is
too high.
Check the function of the fan.
Check the operating temperature.
Nebulizer disconnected
The nebulizer is disconnected
during nebulization. Technical
problem with connection cable.
Connect the nebulizer or change
connection cable.
Nebulizer hardware error
Technical problem with nebulizer
hardware. Temperature too high.
Not enough buffer liquid. Technical
problem with connection cable.
Restart the nebulizer.
Check buffer liquid level. Change
the nebulizer. Change connection
cable. If the problem still remains
contact a service technician.
Nebulizer inhibited due
to overheating
Temperature too high.
Turn off the nebulizer and restart
when cooled down.
Problem
(Display message)
Possible cause
Remedy
Medium priority alarms
Summary of Contents for SERVO-i V3.1
Page 1: ...User s manual US Version VENTILATOR SYSTEM SERVO i V3 1 ...
Page 2: ......
Page 4: ...2 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options ...
Page 16: ...14 Servo User s manual US edition Order No 66 00 261 1 Infant Adult Universal Options Notes ...
Page 71: ...69 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 2 Notes ...
Page 72: ...70 Servo User s manual US edition Order No 66 00 261 2 Infant Adult Universal Options Notes ...
Page 81: ...79 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 3 Notes ...
Page 82: ...80 Servo User s manual US edition Order No 66 00 261 3 Infant Adult Universal Options Notes ...
Page 120: ...118 Servo User s manual US edition Order No 66 00 261 4 Infant Adult Universal Options Notes ...
Page 146: ...144 Servo User s manual US edition Order No 66 00 261 5 Infant Adult Universal Options Notes ...
Page 157: ...155 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 6 Notes ...
Page 158: ...156 Servo User s manual US edition Order No 66 00 261 6 Infant Adult Universal Options Notes ...
Page 191: ...189 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 7 Notes ...
Page 192: ...190 Servo User s manual US edition Order No 66 00 261 7 Infant Adult Universal Options Notes ...
Page 211: ...209 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 8 Notes ...
Page 212: ...210 Servo User s manual US edition Order No 66 00 261 8 Infant Adult Universal Options Notes ...
Page 225: ...223 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 9 Notes ...
Page 226: ...224 Servo User s manual US edition Order No 66 00 261 9 Infant Adult Universal Options Notes ...
Page 242: ...240 Servo User s manual US edition Order No 66 00 261 10 Infant Adult Universal Options Notes ...
Page 256: ...254 Servo User s manual US edition Order No 66 00 261 11 Infant Adult Universal Options Notes ...
Page 259: ...257 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 12 Notes ...
Page 260: ...258 Servo User s manual US edition Order No 66 00 261 12 Infant Adult Universal Options Notes ...
Page 274: ...272 Servo User s manual US edition Order No 66 00 261 13 Infant Adult Universal Options Notes ...
Page 278: ...276 Servo User s manual US edition Order No 66 00 261 14 Infant Adult Universal Options Notes ...
Page 279: ......
Page 283: ...VENTILATOR SYSTEM V 3 1 USER S MANUAL US VERSION SERVO i ...