232
Servo
…
User´s manual
US edition
Order No: 66 00 261
10
Infant
Adult
Universal
Options
O
2
supply pressure: High O
2
supply pressure above 6.5 kPa x
100.
O
2
supply pressure at gas inlet is too
high.
Check the gas supply lines. Perform
a Pre-use check. If the problem still
remains contact a service
technician.
O
2
supply pressure: Low
O
2
supply pressure below 2.0 kPa x
100 or above 6.5 kPa x 100.
O
2
supply pressure at gas inlet is too
low.Gas supply line disconnected.
Note: The O
2
supply too low alarm
can be permanently silenced (Audio
off) when activated.
Check and connect gas supply
lines. Perform a Pre-use check.
Panel disconnected
No communication between User
interface and Patient unit.
Check control cable. If the problem
still remains contact a service
technician.
PEEP High
The measured end expiratory
pressure is above the preset or
default alarm limit for three
consecutive breaths.
Check patient breathing system.
Check patient connection (cuff
pressure/tracheal tube size).
Perform a Pre-use check.
Check ventilator settings.
Check the alarm settings.
PEEP Low
The measured end expiratory
pressure is below the preset or
default alarm limit for three
consecutive breaths.
Note: Setting the alarm to 0 (zero) is
equal to alarm off.
Leakage in patient breathing
system. Leakage at patient
connection (cuff, tracheal tube).
Check patient breathing system.
Check patient connection (cuff
pressure/tracheal tube size).
Perform a Pre-use check.
Check the alarm settings.
Regulation pressure
limited
It is not possible to reach the Set
volume in PRVC and VS, due to
restrictions imposed by the set
Upper pressure limit.
Set high pressure alarm limit, limits
the regulatory pressure used in
PRVC or VS.
Check ventilator settings.
Remove one CO
2
module
Two CO
2
Analyzer modules are
connected at the same time.
Remove one of the CO
2
Analyzer
modules.
Respiratory Rate: High
Respiratory frequency too high.
Auto triggering.
Attend to the patient.
Check the trigger setting.
Respiratory Rate: Low
Respiratory frequency too low.
Trigger sensitivity setting incorrect.
Large tidal volume.
Attend to the patient.
Check trigger setting.
Check Inspiratory cycle-off setting.
Problem
(Display message)
Possible cause
Remedy
Medium priority alarms
Summary of Contents for SERVO-i V3.1
Page 1: ...User s manual US Version VENTILATOR SYSTEM SERVO i V3 1 ...
Page 2: ......
Page 4: ...2 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options ...
Page 16: ...14 Servo User s manual US edition Order No 66 00 261 1 Infant Adult Universal Options Notes ...
Page 71: ...69 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 2 Notes ...
Page 72: ...70 Servo User s manual US edition Order No 66 00 261 2 Infant Adult Universal Options Notes ...
Page 81: ...79 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 3 Notes ...
Page 82: ...80 Servo User s manual US edition Order No 66 00 261 3 Infant Adult Universal Options Notes ...
Page 120: ...118 Servo User s manual US edition Order No 66 00 261 4 Infant Adult Universal Options Notes ...
Page 146: ...144 Servo User s manual US edition Order No 66 00 261 5 Infant Adult Universal Options Notes ...
Page 157: ...155 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 6 Notes ...
Page 158: ...156 Servo User s manual US edition Order No 66 00 261 6 Infant Adult Universal Options Notes ...
Page 191: ...189 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 7 Notes ...
Page 192: ...190 Servo User s manual US edition Order No 66 00 261 7 Infant Adult Universal Options Notes ...
Page 211: ...209 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 8 Notes ...
Page 212: ...210 Servo User s manual US edition Order No 66 00 261 8 Infant Adult Universal Options Notes ...
Page 225: ...223 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 9 Notes ...
Page 226: ...224 Servo User s manual US edition Order No 66 00 261 9 Infant Adult Universal Options Notes ...
Page 242: ...240 Servo User s manual US edition Order No 66 00 261 10 Infant Adult Universal Options Notes ...
Page 256: ...254 Servo User s manual US edition Order No 66 00 261 11 Infant Adult Universal Options Notes ...
Page 259: ...257 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 12 Notes ...
Page 260: ...258 Servo User s manual US edition Order No 66 00 261 12 Infant Adult Universal Options Notes ...
Page 274: ...272 Servo User s manual US edition Order No 66 00 261 13 Infant Adult Universal Options Notes ...
Page 278: ...276 Servo User s manual US edition Order No 66 00 261 14 Infant Adult Universal Options Notes ...
Page 279: ......
Page 283: ...VENTILATOR SYSTEM V 3 1 USER S MANUAL US VERSION SERVO i ...