192
Servo
…
User´s manual
US edition
Order No: 66 00 261
8
Infant
Adult
Universal
Options
The gas that passes into the ventilator´s
inspiration system also passes an inspiratory
filter and is usually clean and dry.
Caution:
• All personnel should be aware of the risk
of parts being infected when
disassembling and cleaning the ventilator.
• All disposable parts must be discarded
according to hospital routine and in an
environmentally safe way.
Important:
• Bacteria from the patient will appear in the
moist environment of the expiration side.
MAQUET recommends the use of Servo
Guard bacteria filter to reduce the
transmission of bacteria from the patient
via the expiratory channel to the Expiratory
cassette. By using the recommended filter
the need for regular cleaning is reduced,
which consequently would result in a
prolonged lifetime of the expiratory
cassette. By regularly replacing the
bacteria filter, the risk of infection being
spread to the staff and the risk of cross
infection between patients are reduced.
• Sterilization is normally not necessary for
the expiratory cassette but when applied
use validated processes only.
• Autoclaving will reduce the lifetime of the
cassette. Lifetime reduction is greater for
autoclaving in prolonged cycles (> 4min) at
134 °C (273 °F).
• Cleaning routines for the Fisher & Paykel
humidifier are described in a separate
operating manual.
• Cleaning routines for the Aeroneb-Pro
(Aeroneb professional Nebulizer system)
are described in a separate operating
manual.
• The single use Y Sensors (Adult/Neonatal)
is supplied NON-STERILE in a plastic bag.
The Y Sensor can be sterilized before use
by EtO or radiation using normal hospital
procedures.
• When handling any part of the Servo-i
Ventilator System, always follow your
hospital‘s guidelines for handling
infectious material. Since cleaning,
sterilization, sanitation, and disinfection
practices vary widely among health care
institutions, it is not possible for MAQUET
to specify on particular practices that will
meet all needs, or to be responsible for the
effectiveness of cleaning, sterilization, and
other procedures carried out in the patient
care setting. MAQUET recommends
methods that have been validated using
the specified equipment and procedures
outlined in this manual. Other methods
may work but are not covered by the
warranty unless MAQUET has given
written permission.
• The result of cleaning/disinfection can be
affected by the quality of water. MAQUET
recommend drinking water as minimum
quality level.
• The most critical step in cleaning/
disinfection is cleaning. It is impossible to
disinfect or even thermally sterilize an
inadequately cleaned instrument. The
reduction of bioburden by cleaning is key
to a good result. Whenever possible
cleaning shall be performed immediately
after use. Foreign particles such as saliva
or blood should not be left to dry on the
devices. Bacteria filters such as Servo
Guard can be used to protect the
Expiratory cassette from contamination, a
practice recommended by MAQUET. If
filters are used and the stated replacement
interval for the filter is observed, the only
cleaning needed is wiping the exterior. If
bacteria filters are not used, MAQUET
recommends using medical dish
disinfectors for cleaning/disinfection.
Hygiene
Summary of Contents for SERVO-i V3.1
Page 1: ...User s manual US Version VENTILATOR SYSTEM SERVO i V3 1 ...
Page 2: ......
Page 4: ...2 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options ...
Page 16: ...14 Servo User s manual US edition Order No 66 00 261 1 Infant Adult Universal Options Notes ...
Page 71: ...69 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 2 Notes ...
Page 72: ...70 Servo User s manual US edition Order No 66 00 261 2 Infant Adult Universal Options Notes ...
Page 81: ...79 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 3 Notes ...
Page 82: ...80 Servo User s manual US edition Order No 66 00 261 3 Infant Adult Universal Options Notes ...
Page 120: ...118 Servo User s manual US edition Order No 66 00 261 4 Infant Adult Universal Options Notes ...
Page 146: ...144 Servo User s manual US edition Order No 66 00 261 5 Infant Adult Universal Options Notes ...
Page 157: ...155 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 6 Notes ...
Page 158: ...156 Servo User s manual US edition Order No 66 00 261 6 Infant Adult Universal Options Notes ...
Page 191: ...189 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 7 Notes ...
Page 192: ...190 Servo User s manual US edition Order No 66 00 261 7 Infant Adult Universal Options Notes ...
Page 211: ...209 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 8 Notes ...
Page 212: ...210 Servo User s manual US edition Order No 66 00 261 8 Infant Adult Universal Options Notes ...
Page 225: ...223 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 9 Notes ...
Page 226: ...224 Servo User s manual US edition Order No 66 00 261 9 Infant Adult Universal Options Notes ...
Page 242: ...240 Servo User s manual US edition Order No 66 00 261 10 Infant Adult Universal Options Notes ...
Page 256: ...254 Servo User s manual US edition Order No 66 00 261 11 Infant Adult Universal Options Notes ...
Page 259: ...257 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 12 Notes ...
Page 260: ...258 Servo User s manual US edition Order No 66 00 261 12 Infant Adult Universal Options Notes ...
Page 274: ...272 Servo User s manual US edition Order No 66 00 261 13 Infant Adult Universal Options Notes ...
Page 278: ...276 Servo User s manual US edition Order No 66 00 261 14 Infant Adult Universal Options Notes ...
Page 279: ......
Page 283: ...VENTILATOR SYSTEM V 3 1 USER S MANUAL US VERSION SERVO i ...