117
Servo
…
User´s manual
US edition
Order No: 66 00 261
Infant
Adult
Universal
Options
4
WARNING! If the trigger sensitivity is set too
high, a self-triggering (auto-triggering)
condition may be reached. This condition
can also be reached if there is leakage in the
breathing system, e.g. if an uncuffed
endotracheal tube is used. Triggering will
then be initiated by the system and not by the
patient.This should always be avoided by
decreasing the trigger sensitivity (page 23).
This is also important during transport as the
movement of the body and the breathing
system may lead to false triggering.
Cautions:
• To prevent the ventilator from tipping over,
the Patient Unit must be pushed into its
locked position during transportation.
• If Battery modules are inserted, the
ventilator should be connected to the
mains when not in use. The Battery
modules are then charged automatically.
• Batteries and O
2
cell must not be
disposed of with ordinary waste.
• Lock the gas cylinders firmly to the Mobile
Cart.
• To prevent the gas cylinders from sliding
out or tipping over, ensure that the straps
are firmly fixed to the center of the
cylinders.
• When moving the Support Arm or
changing position, watch the patient
connection carefully to see that no pulling
or other movement occurs.
• Store the system at a temperature between
-25
o
C and +60
o
C (-13
o
F to 140
o
F) at
<95% RH.
System transport and storage
Summary of Contents for SERVO-i V3.1
Page 1: ...User s manual US Version VENTILATOR SYSTEM SERVO i V3 1 ...
Page 2: ......
Page 4: ...2 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options ...
Page 16: ...14 Servo User s manual US edition Order No 66 00 261 1 Infant Adult Universal Options Notes ...
Page 71: ...69 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 2 Notes ...
Page 72: ...70 Servo User s manual US edition Order No 66 00 261 2 Infant Adult Universal Options Notes ...
Page 81: ...79 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 3 Notes ...
Page 82: ...80 Servo User s manual US edition Order No 66 00 261 3 Infant Adult Universal Options Notes ...
Page 120: ...118 Servo User s manual US edition Order No 66 00 261 4 Infant Adult Universal Options Notes ...
Page 146: ...144 Servo User s manual US edition Order No 66 00 261 5 Infant Adult Universal Options Notes ...
Page 157: ...155 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 6 Notes ...
Page 158: ...156 Servo User s manual US edition Order No 66 00 261 6 Infant Adult Universal Options Notes ...
Page 191: ...189 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 7 Notes ...
Page 192: ...190 Servo User s manual US edition Order No 66 00 261 7 Infant Adult Universal Options Notes ...
Page 211: ...209 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 8 Notes ...
Page 212: ...210 Servo User s manual US edition Order No 66 00 261 8 Infant Adult Universal Options Notes ...
Page 225: ...223 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 9 Notes ...
Page 226: ...224 Servo User s manual US edition Order No 66 00 261 9 Infant Adult Universal Options Notes ...
Page 242: ...240 Servo User s manual US edition Order No 66 00 261 10 Infant Adult Universal Options Notes ...
Page 256: ...254 Servo User s manual US edition Order No 66 00 261 11 Infant Adult Universal Options Notes ...
Page 259: ...257 Servo User s manual US edition Order No 66 00 261 Infant Adult Universal Options 12 Notes ...
Page 260: ...258 Servo User s manual US edition Order No 66 00 261 12 Infant Adult Universal Options Notes ...
Page 274: ...272 Servo User s manual US edition Order No 66 00 261 13 Infant Adult Universal Options Notes ...
Page 278: ...276 Servo User s manual US edition Order No 66 00 261 14 Infant Adult Universal Options Notes ...
Page 279: ......
Page 283: ...VENTILATOR SYSTEM V 3 1 USER S MANUAL US VERSION SERVO i ...