With continuous sampling the measurement data is sent directly to the
computer and Spraytec software for analysis.
.
Timed – as continuous, but the measurement can be requested to stop after
a set amount of time. The stop time is set using the
Minutes:
and
Sec-
onds:
boxes. These boxes only appear when the timed option is selected.
.
Rapid – this is used to measure rapid pulsed spray events, for example nasal
sprays, MDIs or fuel injectors, but can also be used for rapid measurements
of continuous sprays.
The data acquisition rate for the measurement data can be set between
2.5Hz and 2.5kHz. Due to the speed of data acquisition, data obtained
during a rapid measurement is stored by the instrument electronics until
the measurement finishes, then passed to the computer and Spraytec
software for analysis. The Spraytec Measurement Processor downloads the
measurement data from the instrument to the computer.
SOP structure
This graphic shows the basic stages in the
SOP wizard
and the parameters these
are concerned with:
SOP type
– hardware, measurement type and lens.
Documentation
– spray details, operator instructions.
Hardware
– actuator type and details, extraction setup.
Measurement
– background, use of output trigger.
Number of events, trigger type.
Filters, optical, electronic and expiry alarms.
Analysis
– optical properties of the particle
Detectors to use, plume details
Scattering alarms
Curve fit and correlation
Output
– parameters to report, how to average.
When an existing SOP is modified, this tree is shown to allow quick access to any
of the dialogues.
C H A P T E R 5
S
p
r
a
y
t
e
c
P a g e 5 . 6
M
A
N
0
3
6
8
Summary of Contents for Spraytec
Page 1: ...Spraytec User Manual Spraytec User Manual M A N 0 3 6 8 I s s u e 1 0 N o v 2 0 0 5...
Page 7: ...Part 1 Operator s guide Part 1 Operator s guide...
Page 8: ......
Page 9: ...Introduction to this manual Introduction to this manual C H A P T E R 1...
Page 10: ......
Page 16: ...C H A P T E R 1 S p r a y t e c Page 1 6 M A N 0 3 6 8...
Page 17: ...What is the Spraytec What is the Spraytec C H A P T E R 2...
Page 18: ......
Page 31: ...The Spraytec hardware The Spraytec hardware C H A P T E R 3...
Page 32: ......
Page 42: ...C H A P T E R 3 S p r a y t e c Page 3 10 M A N 0 3 6 8...
Page 43: ...The Spraytec software The Spraytec software C H A P T E R 4...
Page 44: ......
Page 63: ...Measurement tutorial Measurement tutorial C H A P T E R 5...
Page 64: ......
Page 99: ...Interpreting the results Interpreting the results C H A P T E R 6...
Page 100: ......
Page 125: ...Using measurement files Using measurement files C H A P T E R 7...
Page 126: ......
Page 133: ...Basic maintenance Basic maintenance C H A P T E R 8...
Page 134: ......
Page 139: ...Part 2 Supervisor s guide Part 2 Supervisor s guide...
Page 140: ......
Page 141: ...Security Security C H A P T E R 9...
Page 142: ......
Page 151: ...Advanced features Advanced features C H A P T E R 1 0...
Page 152: ......
Page 170: ...C H A P T E R 1 0 S p r a y t e c Page 10 18 M A N 0 3 6 8...
Page 171: ...SOP management SOP management C H A P T E R 1 1...
Page 172: ......
Page 175: ...Part 3 Appendices Part 3 Appendices...
Page 176: ......
Page 177: ...Health and Safety Health and Safety A P P E N D I X A...
Page 178: ......
Page 184: ...A P P E N D I X A S p r a y t e c Page A 6 M A N 0 3 6 8...
Page 185: ...Specification Specification A P P E N D I X B...
Page 186: ......
Page 191: ...Site requirements f r o m M A N 0 3 6 8 I s s u e 1 0 Site requirements A P P E N D I X C...
Page 200: ...A P P E N D I X D S p r a y t e c Page D 3 M A N 0 3 6 8...
Page 201: ...Installation Installation A P P E N D I X E...
Page 202: ......
Page 206: ...A P P E N D I X E S p r a y t e c Page E 4 M A N 0 3 6 8...
Page 207: ...Regulatory statements Regulatory statements A P P E N D I X F...
Page 208: ......
Page 213: ...Index Index...
Page 214: ......
Page 221: ......