After displaying the electrical background, the optical background is shown.
This is shown as a percentage value. Typically on a clean instrument the
value should be between 70% and 95%, accompanied by a background of
less than 50 for a dry measurement, or 100 for a wet measurement.
If the laser intensity is low (around 60%) and the background level is over
200 the instrument optics need cleaning. If the optics have been cleaned to
give an acceptable background, but the laser intensity is still low (60% or
below) the laser may be failing; contact Malvern Instruments.
.
If an
Inspection Stage
was specified by the SOP, the
Inspection
tab
shown below appears. Spray some of the material into the measurement
zone and observe the obscuration and light energy histogram the dialogue
shows. These should be enough to confirm whether the instrument
settings and any accessories are set up correctly:
.
If there is anything obviously wrong, it’s possible to change the setup here.
Otherwise, click the
Start
button to begin the measurement. Follow any
instructions given.
.
Now the checks have been performed, the
Measure sample
tab appears as
the third tab in the dialogue and the
Start
button becomes available. (The
first two tabs in the window remain available for repeating background or
alignment checking if the user requires).
C H A P T E R 5
S
p
r
a
y
t
e
c
P a g e 5 . 3 2
M
A
N
0
3
6
8
Summary of Contents for Spraytec
Page 1: ...Spraytec User Manual Spraytec User Manual M A N 0 3 6 8 I s s u e 1 0 N o v 2 0 0 5...
Page 7: ...Part 1 Operator s guide Part 1 Operator s guide...
Page 8: ......
Page 9: ...Introduction to this manual Introduction to this manual C H A P T E R 1...
Page 10: ......
Page 16: ...C H A P T E R 1 S p r a y t e c Page 1 6 M A N 0 3 6 8...
Page 17: ...What is the Spraytec What is the Spraytec C H A P T E R 2...
Page 18: ......
Page 31: ...The Spraytec hardware The Spraytec hardware C H A P T E R 3...
Page 32: ......
Page 42: ...C H A P T E R 3 S p r a y t e c Page 3 10 M A N 0 3 6 8...
Page 43: ...The Spraytec software The Spraytec software C H A P T E R 4...
Page 44: ......
Page 63: ...Measurement tutorial Measurement tutorial C H A P T E R 5...
Page 64: ......
Page 99: ...Interpreting the results Interpreting the results C H A P T E R 6...
Page 100: ......
Page 125: ...Using measurement files Using measurement files C H A P T E R 7...
Page 126: ......
Page 133: ...Basic maintenance Basic maintenance C H A P T E R 8...
Page 134: ......
Page 139: ...Part 2 Supervisor s guide Part 2 Supervisor s guide...
Page 140: ......
Page 141: ...Security Security C H A P T E R 9...
Page 142: ......
Page 151: ...Advanced features Advanced features C H A P T E R 1 0...
Page 152: ......
Page 170: ...C H A P T E R 1 0 S p r a y t e c Page 10 18 M A N 0 3 6 8...
Page 171: ...SOP management SOP management C H A P T E R 1 1...
Page 172: ......
Page 175: ...Part 3 Appendices Part 3 Appendices...
Page 176: ......
Page 177: ...Health and Safety Health and Safety A P P E N D I X A...
Page 178: ......
Page 184: ...A P P E N D I X A S p r a y t e c Page A 6 M A N 0 3 6 8...
Page 185: ...Specification Specification A P P E N D I X B...
Page 186: ......
Page 191: ...Site requirements f r o m M A N 0 3 6 8 I s s u e 1 0 Site requirements A P P E N D I X C...
Page 200: ...A P P E N D I X D S p r a y t e c Page D 3 M A N 0 3 6 8...
Page 201: ...Installation Installation A P P E N D I X E...
Page 202: ......
Page 206: ...A P P E N D I X E S p r a y t e c Page E 4 M A N 0 3 6 8...
Page 207: ...Regulatory statements Regulatory statements A P P E N D I X F...
Page 208: ......
Page 213: ...Index Index...
Page 214: ......
Page 221: ......