Externally-triggered spray SOP
This section describes how to use the SOP wizard to create an SOP for use with
an externally-triggered spray, for example a fuel spray.
.
In the
Welcome
dialogue select the
Default rapid
template and click
Next
. The
SOP Type
dialogue is displayed:
Click the
Hardware configuration
button and select
Open bench
with accessories
. Keep the
Basic settings
measurement type
Rapid
and
the maximum
Data acquisition rate
(2.5 kHz). Click
Next
.
'
N o t e
.
The
Lens type
can only be changed on dual lens systems. Change to
the appropriate lens if required(also see Chapter 8).
.
In the
Labels
dialogue specify any required information on the spray, its
source and lot, etc. to allow the spray to be traced in the future. If required,
the operator can change/add to these when they run the test. Click
Next
.
.
In the
Instructions
dialogue type any instructions for operators to follow
at the SOP start or end. For either, select the relevant check box and type in
the instruction text below it. These instructions will pop up on the screen
when an operator runs the SOP. Click
Next
.
.
In the
Actuator
dialogue select the actuator to use from those listed. A set
of parameters for the actuator appears; set these according to the manufac-
turer’s documentation.
C H A P T E R 5
S
p
r
a
y
t
e
c
P a g e 5 . 2 3
Summary of Contents for Spraytec
Page 1: ...Spraytec User Manual Spraytec User Manual M A N 0 3 6 8 I s s u e 1 0 N o v 2 0 0 5...
Page 7: ...Part 1 Operator s guide Part 1 Operator s guide...
Page 8: ......
Page 9: ...Introduction to this manual Introduction to this manual C H A P T E R 1...
Page 10: ......
Page 16: ...C H A P T E R 1 S p r a y t e c Page 1 6 M A N 0 3 6 8...
Page 17: ...What is the Spraytec What is the Spraytec C H A P T E R 2...
Page 18: ......
Page 31: ...The Spraytec hardware The Spraytec hardware C H A P T E R 3...
Page 32: ......
Page 42: ...C H A P T E R 3 S p r a y t e c Page 3 10 M A N 0 3 6 8...
Page 43: ...The Spraytec software The Spraytec software C H A P T E R 4...
Page 44: ......
Page 63: ...Measurement tutorial Measurement tutorial C H A P T E R 5...
Page 64: ......
Page 99: ...Interpreting the results Interpreting the results C H A P T E R 6...
Page 100: ......
Page 125: ...Using measurement files Using measurement files C H A P T E R 7...
Page 126: ......
Page 133: ...Basic maintenance Basic maintenance C H A P T E R 8...
Page 134: ......
Page 139: ...Part 2 Supervisor s guide Part 2 Supervisor s guide...
Page 140: ......
Page 141: ...Security Security C H A P T E R 9...
Page 142: ......
Page 151: ...Advanced features Advanced features C H A P T E R 1 0...
Page 152: ......
Page 170: ...C H A P T E R 1 0 S p r a y t e c Page 10 18 M A N 0 3 6 8...
Page 171: ...SOP management SOP management C H A P T E R 1 1...
Page 172: ......
Page 175: ...Part 3 Appendices Part 3 Appendices...
Page 176: ......
Page 177: ...Health and Safety Health and Safety A P P E N D I X A...
Page 178: ......
Page 184: ...A P P E N D I X A S p r a y t e c Page A 6 M A N 0 3 6 8...
Page 185: ...Specification Specification A P P E N D I X B...
Page 186: ......
Page 191: ...Site requirements f r o m M A N 0 3 6 8 I s s u e 1 0 Site requirements A P P E N D I X C...
Page 200: ...A P P E N D I X D S p r a y t e c Page D 3 M A N 0 3 6 8...
Page 201: ...Installation Installation A P P E N D I X E...
Page 202: ......
Page 206: ...A P P E N D I X E S p r a y t e c Page E 4 M A N 0 3 6 8...
Page 207: ...Regulatory statements Regulatory statements A P P E N D I X F...
Page 208: ......
Page 213: ...Index Index...
Page 214: ......
Page 221: ......