The Magstim Company Ltd.
©
73
NOP08-EN-10
5.5 Neurosign V4 Earthing Cable
In cases of electrical interference, the earthing cable (PN: 210230) should be fitted
to the Neurosign V4 equipotential point, found near the power switch, and to the
hospital equipotential wall socket (POAG socket) as below.
Problem
Cause
Action
Pre-amplifier offline/
lost connection
The pre-amplifier is either
not connected, or there
is a problem with the
connection.
If you are not currently monitoring you
can continue with monitoring disabled by
pressing 'YES' when prompted.
Alternatively, disconnect the pre-amplifier
lead from the pre-amplifier connection
socket on the rear of the screen.
Reconnect the pre-amplifier, if the error
remains contact your service provider.
Contact details can be found on page
Summary of Contents for NEUROSIGN V4
Page 4: ...The Magstim Company Ltd iv NOP08 EN 10 Page intentionally left blank...
Page 6: ...The Magstim Company Ltd vi NOP08 EN 10 Page intentionally left blank...
Page 8: ...The Magstim Company Ltd viii NOP08 EN 10 Page intentionally left blank...
Page 10: ...The Magstim Company Ltd 10 NOP08 EN 10 Page intentionally left blank...
Page 12: ...The Magstim Company Ltd 12 NOP08 EN 10 Page intentionally left blank...
Page 13: ...The Magstim Company Ltd 13 NOP08 EN 10 SECTION 1 PRODUCT OVERVIEW...
Page 18: ...The Magstim Company Ltd 18 NOP08 EN 10 Page intentionally left blank...
Page 19: ...The Magstim Company Ltd 19 NOP08 EN 10 SECTION 2 ACCESSORIES...
Page 25: ...The Magstim Company Ltd 25 NOP08 EN 10 SECTION 3 WARNINGS SYMBOLS...
Page 35: ...The Magstim Company Ltd 35 NOP08 EN 10 SECTION 4 INSTRUCTIONS FOR USE...
Page 65: ...The Magstim Company Ltd 65 NOP08 EN 10 SECTION 5 TROUBLE SHOOTING...
Page 74: ...The Magstim Company Ltd 74 NOP08 EN 10 Page intentionally left blank...
Page 75: ...The Magstim Company Ltd 75 NOP08 EN 10 SECTION 6 CARE MAINTENANCE...
Page 79: ...The Magstim Company Ltd 79 NOP08 EN 10 SECTION 7 TECHNICAL INFORMATION...