• Any attempt to remove the security screws will invalidate the product guarantee.
• No unauthorised modifications may be made to the nerve monitor or any of its
associated accessories.
• CLASS I: Equipment must only be connected to an AC mains supply with protective
earth. Connection should only be made with the power lead provided with the
equipment.
• This device is designed as an aid to surgeons in the identification and monitoring
of nerves at risk of injury during surgery. It is not a replacement for surgical
experience.
• Do not use this equipment in the presence of flammable anaesthetic mixtures.
• Handle with care. Dropping this device or any accessories may cause physical
damage and affect performance.
• Ensure that the mains plug and socket are readily accessible to provide isolation
from the supply mains if necessary.
• Do not permit liquid close to the ventilation slots in the bottom of the casing.
• Do not insert objects into connector housings, holes or slots in the casing for
cleaning purposes, as this may cause damage.
• In the event of hardware failure or software corruption the nerve monitor is
designed to fail in a safe state. In this state the stimulator and EMG audio channels
are disabled.
• Do not connect the mute sensor to the nerve monitor without attaching the clamp
to the electrocautery cable or spurious muting of the nerve monitor may occur.
• When a mute sensor is connected and active, the stimulator current confirm will
not be reported.
• High levels of radio frequency (RF) may interfere with the operation of the mute
sensor if the device is not connected to the electrocautery cable.
• Care should be taken to avoid the tip of the stimulating probe coming into contact
with that of the electrocautery probe, this will cause damage to the internal
circuitry of the nerve monitor.
• The use of this device other than as described will reduce the level of protection
provided by the equipment. If it is deemed necessary to use the device in any
way other than as stated in this document it is strongly advised to contact The
Magstim Company Ltd. for advice before use.
• During use, the nerve monitor will have a number of leads/cables connected.
Therefore, care should be exercised with lead/cable placement to avoid a tripping
hazard.
Summary of Contents for NEUROSIGN V4
Page 4: ...The Magstim Company Ltd iv NOP08 EN 10 Page intentionally left blank...
Page 6: ...The Magstim Company Ltd vi NOP08 EN 10 Page intentionally left blank...
Page 8: ...The Magstim Company Ltd viii NOP08 EN 10 Page intentionally left blank...
Page 10: ...The Magstim Company Ltd 10 NOP08 EN 10 Page intentionally left blank...
Page 12: ...The Magstim Company Ltd 12 NOP08 EN 10 Page intentionally left blank...
Page 13: ...The Magstim Company Ltd 13 NOP08 EN 10 SECTION 1 PRODUCT OVERVIEW...
Page 18: ...The Magstim Company Ltd 18 NOP08 EN 10 Page intentionally left blank...
Page 19: ...The Magstim Company Ltd 19 NOP08 EN 10 SECTION 2 ACCESSORIES...
Page 25: ...The Magstim Company Ltd 25 NOP08 EN 10 SECTION 3 WARNINGS SYMBOLS...
Page 35: ...The Magstim Company Ltd 35 NOP08 EN 10 SECTION 4 INSTRUCTIONS FOR USE...
Page 65: ...The Magstim Company Ltd 65 NOP08 EN 10 SECTION 5 TROUBLE SHOOTING...
Page 74: ...The Magstim Company Ltd 74 NOP08 EN 10 Page intentionally left blank...
Page 75: ...The Magstim Company Ltd 75 NOP08 EN 10 SECTION 6 CARE MAINTENANCE...
Page 79: ...The Magstim Company Ltd 79 NOP08 EN 10 SECTION 7 TECHNICAL INFORMATION...