background image

M3LPA2

P. 8 / 10

EM-2657 Rev.3

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN

Phone: +81(6)6659-8201   Fax: +81(6)6659-8510   E-mail: [email protected]

 I/O RANGE LABEL

Blank I/O range labels are included in the product pack-

age.  Write in the configured ranges and put the label on 

the side as shown below.

INPUT
OUTPUT

I/O Range Label (included in the product package)

 OUTPUT TYPE & RANGE 

Table 14

OUTPUT TYPE 

MINIMUM SPAN 

MAXIMUM RANGE 

CONFORMANCE RANGE

DC Current 

1mA 

0 to 20mA 

0 to 24mA 

 

DC Voltage, Narrow Spans 

250mV 

-2.5 to +2.5V 

-3 to +3V 

 

DC Voltage, Wide Spans 

1V 

-10 to +10V 

-11.5 to +11.5V

 INPUT TYPE & RANGE 

Table 13

INPUT TYPE 

MINIMUM AMPLITUDE 

MAXIMUM RANGE  

MAXIMUM AMPLITUDE

Open Collector 

4V 

0 – 200 kHz 

12V 

Mechanical Contact 

4V 

0 – 10 Hz 

12V 

 

Voltage Pulse 

0.1V 

0 – 200 kHz 

100V

Two-wire Current Pulse 

4mA 

0 – 200 kHz 

25mA 

 

RS-422 Line Driver Pulse 

–– 

0 – 200 kHz 

––

 

 ZERO & SPAN ADJUSTMENTS

After the transmitter is installed and operational, fine zero 

and span tuning can be performed as explained below.  Both 

zero and span are adjustable within ±15%.
1)  Run  Mode:  Confirm  that  the  green  LED  is  blinking 

(model M3LPA2-x/A) or the green LED turns on (model 

M3LPA2-x/B).

2)  Fine Zero Calibration Mode: Press MODE button for 1 

or 2 seconds until the LD1 red light is ON and the LD2 

green light is blinking.

3)  Use UP (increase) and DOWN (decrease) buttons to ad

-

just the output to 0%.

4)  Fine  Span  Calibration  Mode:  Push  MODE  button  and 

confirm that the LD1 red light is ON and the LD3 green 
light instead of LD2 is blinking.

5)  Use UP (increase) and DOWN (decrease) buttons to ad

-

just the output to 100%.

6)  Run Mode: Push MODE button and confirm that only the 

LD1 green light is blinking.

Note 1: Calibration steps can be skipped when not needed 

by repeating pushing MODE buttons.

Note 2: There is no limitation of entering values for multi-
ple times within one step of Calibration Mode.  Signal level 

is stored each time the respective UP or DOWN button is 
pressed.  However, for the input ranging, end always with 
100% (UP).
Note 3: Once the transmitter enters the fine zero or span 
calibration mode, the output goes to preadjusted 0% or 100% 

value respectively regardless of the present input signal.

Summary of Contents for M3LPA2

Page 1: ...unted inside a panel Insert noise filters for the power source connected to the unit Densei Lambda Model MBS 1220 22 or equivalent is recommended The actual installation environments such as panel con...

Page 2: ...insert a mi nus driver between the euro type connector terminal block and the housing body pull up the driver and pull out the euro type connector terminal block INSTALLATION DIN RAIL MOUNTING Set th...

Page 3: ...NPUT External DC Supply DC POWERED TYPE 2 Be aware that the AC power and DC power connect to different terminals 2 DC POWER INPUT 1 U V 1 3 2 6 5 4 CONFIGURATOR JACK 7 8 10 12 OUTPUT AC POWERED TYPE 2...

Page 4: ...d the pulse sensing type and noise filter must be selected with the DIP SW4 1 through SW4 3 See Tables 9 and 10 For detailed information on the PC configuration refer to the M3LPACFG instruction manua...

Page 5: ...h the large noise filter setting and to 10 msec with the small noise filter setting OUTPUT TYPE SW3 1 Table 11 OUTPUT SW3 3 SW3 2 SW1 4 SW1 3 SW1 2 SW1 1 0 20mA OFF ON OFF ON OFF OFF 2 5 2 5V ON OFF O...

Page 6: ...ght is ON and the LD2 red light is blinking 5 0 Input Ranging Apply the desired minimum input level e g 10 Hz from the simulator and push DOWN button until the LD1 flashes for approx 5 sec and then tu...

Page 7: ...A the green LED is blinking For the model M3LPA2 x B the green LED turns on In order to change the threshold apply simulated 100 input signal and push UP increase or DOWN decrease buttons until the ou...

Page 8: ...explained below Both zero and span are adjustable within 15 1 Run Mode Confirm that the green LED is blinking model M3LPA2 x A or the green LED turns on model M3LPA2 x B 2 Fine Zero Calibration Mode...

Page 9: ...VITY SCALE 50 100V p p 1 20 25 50V p p 1 10 10 25V p p 1 5 5 10V p p 1 2 1 5V p p 1 0 5 1V p p 1 0 1 0 5V p p 1 A specific sensitivity scale is applied according to the pulse amplitude by setting up t...

Page 10: ...operating conditions DIP SW Configuration RUN Mode The transmitter is configured via DIP SW and is in normal operating conditions Output Saturated in PC Configuration Mode Option A The output is satur...

Reviews: