47DV
P. 13 / 34
EM-9501-B Rev.6
■
OPERATIONS IN MEASURING MODE
• Switching the main display to MAX or MIN values
Press Max/Min button to switch the main display to MAX or MIN values.
‘Max’ or ‘Min’ indicator turns on during the MAX/MIN display mode.
Press Max/Min button again for 1 second or more, or turn the power supply off and on to cancel the MAX/MIN display mode.
MAX or MIN values are not updated while in the stand-by sequence.
MAX or MIN values since the startup or the resetting are valid during an event triggered measuring.
• Forced Zero
Press Up button to shift the present display value to zero and to continue measuring in reference to this point.
‘FZ’ indicator turns on during the Forced Zero mode.
Press Up button again for 1 second or more to cancel the Forced Zero mode.
Forced Zero cannot be controlled when the operation is prohibited with the lockout protection setting (LV2).
Forced Zero cannot be executed while in the MAX/MIN display mode.
Forced Zero value is stored in memory even when the power is removed.
Caution: Forced Zero may be unintentionally executed depending on the timing of your pressing Up button with Alarm/ but-
ton when entering Infrared Communication Mode.
Max. Value
Display
Min. Value
Display
Max/Min
button
Max/Min
button
Max/Min
button
ON
Measuring
Mode
Max
ON
Min
Forced Zero
Display
Up button
Up button
1 second or longer
Measuring
Mode
ON
FZ
Display
Forced Zero
Canceled
Time
Up button
1 sec
Forced Zero
Executed
Measuring values after Forced Zero