220
47DM OPERATING MANUAL EM-9510-B Rev.6
Alarm status indication
All setpoints can be set and indicated regardless of alarm output options. Each is
independently set either for Hi or Lo alarm trip.
LL indicator
Turns on in red when the LL alarm is tripped
L indicator
Turns on in red when the L alarm is tripped
H indicator
Turns on in green when the H alarm is tripped
HH indicator
Turns on in green when the HH alarm is tripped
P indicator
Turns on in amber when none of the other alarms is tripped
Status indicators
Max, Min, FZ, TZ
Display MAX/MIN value, amber LED turns on at Forced Zero mode and Tare
Adjustment mode
Function indicators
Hld
Turns on in green when HOLD signal is ON
TG
Turns on in green when TIMING signal is ON
NG
Blinking in green when a parameter is invalid
Zro
Turns on in green at zero setting of scaling setting mode
Spn
Turns on in green at span setting of scaling setting mode
Tch
Turns on in green at input scaling, blinking in red at teach calibration
■
EXCITATION SUPPLY
+12 V sensor excitation Output voltage (across
the terminals 5 – 6)
12 – 16 V DC with no load
10.8 V DC minimum at 80 mA
Current rating
≤ 84 mA DC
Shortcircuit Protection
Current limited
97 mA maximum
Protected time duration
No limit
+24 V 2-wire transmit-
ter excitation
Output voltage (across
the terminals 5 – 6)
24 – 28 V DC with no load
22 V DC minimum at 20 mA
Current rating
≤ 22 mA DC
Shortcircuit Protection
Current limited
30 mA maximum
Protected time duration
No limit
■
INPUT SPECIFICATIONS
Total resistance
100 Ω – 10 kΩ
Minimum span
10% of total resistance
Excitation
0.5 V DC
■
DC OUTPUT SIGNAL SPECIFICATIONS
DC voltage
Analog output type:
0-5V
Output range
0 – 5 V
Operational range
-0.5 – +5.5 V
Load resistance
2000 Ω minimum
Analog output type: 5V
Output range
±5 V
Operational range
-6 – +6 V
Load resistance
4000 Ω minimum
Analog output type: 10V
Output range
±10 V
Operational range
-12 – +12 V
Load resistance
8000 Ω minimum
DC current
Analog output type:
0-20MA
Output range
0 – 20 mA
Operational range
-2 – +22 mA
Load resistance
400 Ω maximum
Analog output type:
4-20MA
Output range
4 – 20 mA
Operational range
2.4 – 21.6 mA
Load resistance
400 Ω maximum