47DM OPERATING MANUAL EM-9510-B Rev.6
4
Apply 0% input and press Up button to register the value.
• The teach calibration (Zero) is registered.
• ‘Tch’ indicator turns on in green.
IMPORTANT
• Confirm that the input signal is stable before pressing Up button.
• Set ‘input scaling value Zero + 10% of total resistance ≤ input scaling value Span’.
5
■
TO GO ON TO SET THE DISPLAY SCALING VALUE ZERO,
Skip to Step 3 in “3.2 STEP 2. DISPLAY SCALING VALUE ZERO”.
■
TO QUIT,
Hold down Alarm/
↓
or Scale/
↑
button for 1 second or more to return to Measuring Mode.
NOTE
■
IF THE FRONT BUTTONS ARE LEFT UNTOUCHED…
• The display goes back automatically to Measuring Mode after the specified time period (default: 15 sec.).
• This time period (automatic return time) is configurable. (Refer to 14. GOING BACK AUTOMATICALLY TO MEASUR-
■
TO ABORT A SETTING…
• Hold down Max/Min button for 1 second or more in Step 3 to return to Measuring Mode without applying the last
changes.
• If you get lost in a setting mode, you can execute initialization. (Refer to 34.2 INITIALIZING SETTING VALUES.)