RCT 5023-G Reference Manual Rev 1.2
Page 8 of 57
Installation
If this module was supplied as part of a system it is already installed in the rack
enclosure. If the module was supplied as a field upgrade please follow the installation
procedure below.
NOTE Observe static precautions when handling card. Please see ESD
warnings on Page 5.
We recommend you power the rack down before installing any additional modules into an
existing card frame although the RCT 5023-G supports hot swapping.
1. Slide the card module into the card frame. The card should fit easily and should
not require excessive force to insert - if you feel any resistance, there could be
something wrong with the rear connection panel location.
Do not
try and force
the connection this may damage the connectors.
2. The R FR 5012/5014 Card frame has a slot configured for the RCT 5023-G
Controller and the card is simply plugged in where shown.
3.
Power up the rack and check the module LED’s. Check the module is
automatically logged into the control system device tree. (It may take a few
seconds for the control system to “discover” the new module)
NOTE
.
The use of the optional control system is mandatory for the control and
setup of this module. If you do not have the control system, then please contact
your LYNX representative for details on how to upgrade your rack with the LYNX
control system.
RCT 5023-G Controller Card