background image

User Manual for CLX Xenon Light Source

ENGLISH

Summary of Contents for 00-CLX

Page 1: ...User Manual for CLX Xenon Light Source ENGLISH...

Page 2: ...eshooting 9 Maintenance and Cleaning 10 Specifications 10 Electromagnetic Compatibility EMC User Information 11 14 Product Warranty 15 Repair and Return 15 TABLE OF SYMBOLS Found on Light Source Atten...

Page 3: ...and the light cable can cause eye damage if viewed directly for more than momentary periods of time Do not look directly into the light source or the cable Do not point the light cable at another pers...

Page 4: ...The CLX light source comes with a hospital grade power cord Please verify that both the light source and the power cord have been received undamaged All fiber optic light guides should be properly cle...

Page 5: ...t occur when the standby switch is used to control the lamp Note that if the light source is left in standby with power applied with the main power on off switch in the on position and the on standby...

Page 6: ...nsure against failure Figure 7 Figure 6 Retaining clips locked Replace lamp when illumination is no longer satisfactory or sufficient for use To remove the lamp 1 Make sure the main power on off switc...

Page 7: ...main power on off switch to the on position then press on standby switch to turn the unit on 5 Adjust attenuator ring to verify that the unit emits light Figure 9 Figure 8 Forward Lamp module Lamp dis...

Page 8: ...dle screw positioned at the top of the handle slide the handle onto the upper column so there is about 5 inches 13 cm of space between the handle and the top of the column Tighten handle screw to hold...

Page 9: ...e 850358 User Manual for the CLX Adapter Options 401193 Storz adapter Lamp 00 1187 CLX Lamp replacement module Power Cords 601949 AUSTL6M Australian Power Cord 6m 601949 EUR6M European Power Cord 6m 6...

Page 10: ...air the unit if the lamp fails in use Turn off the unit and allow it to cool for at least 15 minutes then try to restart the unit If the lamp still fails to illuminate it may need to be replaced If yo...

Page 11: ...nominal AC Power Leakage Leakage current to chassis with ground wire intact less than 100 microamps Leakage current to chassis with ground wire interrupted less than 500 microamps Classification Type...

Page 12: ...ironment of use to permit the CLX Light Source to perform its intended use without disturbing other Equipment and Systems or non medical electrical equipment Guidance and Manufacturer s Declaration Em...

Page 13: ...commercial or hospital environment Surge IEC 61000 4 5 1kV Differential 2kV Common 1kV Differential 2kV Common Mains power quality should be that of a typical commercial or hospital environment Volta...

Page 14: ...vironment specified below The customer or user of the CLX Light Source should ensure that it is used in such an environment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environm...

Page 15: ...or user of the CLX Light Source can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the CLX Light Source as reco...

Page 16: ...uxtec subjected to unusual physical or electrical stress altered with non Luxtec parts or damaged during shipment back to Luxtec This warranty is non transferable unless authorized in writing by Luxte...

Page 17: ...the Essential Requirements specified in Annex 1 of the Medical Device Directive MDD 99 Hartwell Street West Boylston MA 01583 USA Telephone 800 325 8966 USA only or 1 508 835 9700 Facsimile 1 508 835...

Reviews: