63
particular, the apparatus furtherincludes a phase compensation processing means for receiving from
the subject organisms of biological signals and the driver circuit to provide the amplified signal, the
output is sent to a feedback signal by measuring biometric.
9.
HR ALM SETUP: setup heart rate hight limit ,middle limit ,low limit of the upper and lower limits .
10.
ST AMALYSIS : ST Segment Analysis Select this item and enter the ―ST Segment Analysis‖ menu.
11.
ARR ANALYSIS: ARR Analysis Select this item and enter the ―ARR Analysis‖ menu.
12.
ECG CAL : When the ECG calibration, you can not monitor the patient . Tip: in the middle of the
screen of the instrument calibration can not monitor the patient . Stop the calibration is required to
return to the ―ECG Setup‖ select to ―stop ECG calibration‖ menus .
13.
DEFAULT : Select "default configuration" dialog box, the user can choose "No" or ―Yes" to be ―the
default configuration‖ or ―original configuration will be set‖.
11.5.4
About ST Monitoring
The monitor performs ST segment analysis on normal and atrially paced beats and calculates ST
segment elevations and depressions. This information can be displayed in the form of ST numerics and
snippets on the monitor.
All available leads can be monitored continuously. The ECG waveform does not need to be displayed on
the Screen for ST segment analysis.
Warning
This monitor provides ST level change information; the clinical significance of the ST level
change information should be determined by a physician.
11.5.4.1
Setting of ST Segment Analysis Menu
ST Segment Analysis: this switch is mainly used to set the state of ST Segment Analysis. Only when the
switch is ON, ST Segment Analysis can proceed.
Alarm Switch: Where ―ON‖ is selected, the alarm prompt and saving will proceed when ST analysis result
is alarmed; where ―OFF‖ is selected, alarming will not happen, but
will be prompted beside ST in the
screen parameter area. ST Alarm will be triggered only when its measured value exceeds ST Alarm Upper
Limit or ST Alarm Lower Limit.
Alarm Level: used to set the ST Alarm Level according to three options such as ―High‖, ―Middle‖ and
―Low‖.
Alarm Record: when it is set at ―ON‖, the system will start the recorder for alarm record.
Alarm Upper Limit: used to set the alarm upper limit of ST Segment, of which maximum upper limit is 2.0
and of which minimum upper limit must be more than -1.6 of the set lower limit.
Alarm Lower Limit: used to set the alarm lower limit of ST Segment, of which minimum lower limit is 1.8
and of which maximum lower limit must be lower than -2 of the set upper limit.