- 153 -
(
10
)
Respiration specification
Name
Specifications
Way
Thoracic impedance method
Respiration rate detection
and accuracy
Measuring range
Big
6bpm-120bpm
Small
6bpm-150bpm
Accuracy
±1bpm
Accuracy and error of
preset alarm of respiration
rate
Big
Upper
limit
not
narrower than
8bpm-120bpm
Lower
limit
not
narrower than
6bpm-118bpm
Small
Upper
limit
not
narrower than
8pm-150bpm
Lower
limit
not
narrower than
6bpm-148bpm
Error
±1bpm
Range and error
of
suffocation alarm
Scope:
10s
~
40s
,
and no alarm
(
11
)
CO
2
specification
Name
Specifications
CO
2
‘s
range
of
measurement
0mmHg
~
150mmHg, And air pressure is supplied by the mainframe.
CO
2
‘s resolution
1mmHg or 0.1kPa or 0.1%
Gain
×0 .25 、×0.5、×1、×2、×4
Unit
mmHg、KPa
CO
2
‘s
accuracy
±2mmHg between 0mmHg
~
40mmHg
;
±5% between
41mmHg
~
70mmHg;
±8% between 71mmHg
~
100mmHg;
±10% between 101mmHg
~
150mmHg
Atmospheric
pressure/altidude
400
~
850mmHg/0
~
5029.2 M
O2 compensation
5
~
100 mmHg
Range of alarm set and
error of alarm set
0~ 150mmHg/0
~
20 KPa
;
1mmHg/
0.1 KPa
。
Wave speed
2.5mm/s、25 mm/s
Equilibrium gas
N2O,indoor air ,Helium
(
12
)
IBP Specifications
Name
Specifications
Number of IBP channel
2 channels
Names of channel
pressure
Artery pressure(ART), pulmonary artery pressure(PA), central vein
pressure(CVP), right atrial pressure(RAP), left atrial pressure(LAP),
intracranial pressure(ICP), plus pressure(P1,P2).
Range and accuracy of
measurement
ART
0
~
40kPa
(
0
~
300mmHg
)
PA
-0.8
~
16kPa(-6
~
120 mmHg)
CVP
-1.3
~
5.3kPa(-10
~
40mmHg)
RAP
-1.3
~
5.3kPa(-10
~
40mmHg
)