34
Prompting message means the related information will appear on the physiological alarm area or technical
alarm area of the monitor when the alarm is active. The system will show different alarm levels in different
background colors:
High level alarm: Red
Medium level alarm: Yellow
Low level alarm: Yellow
Levels of the prompt message in the front of the alarm will be distinguished with the following symbols:
High level alarm: ***
Medium level alarm: **
Low level alarm: *
5.7
Alarm Parameter Flashing
When a parameter is alarming, the parameter will flash once every second.
5.8
Icon of Alarm Status
Besides the alarm modes mentioned above, the following alarm icons will appear on the screen to indicate
different alarm conditions.
: indicating alarm muted.
: indicating shut off of a parameter alarm.
5.9
Setting the upper and lower limits of alarm
Take ECG for example:
a)
Select ECG in the parameter area;
b)
In the popped-out menu ECG Setting, select the HR ALM SETUP;
c)
Set the upper and lower limits of HIGH LIMIT as 120bpm and 50bpm respectively;