103
Reminders
Causes
Alarming levels
T over-high
TEMP measurement value higher than the set upper limit
User defined
T over-low
TEMP measurement value lower than the set lower limit
User defined
Technical alarms
Reminders
Causes
Alarming
levels
Solutions
Sensor
disconnected
TEMP cable disconnected
from the monitor
Low
Make sure the cable is reliably
connected
Alarming limit
error
Function/ safety failure
High
Stop using the TEMP alarm function
and inform biochemical engineers or
our maintenance personnel
Reminders:
Reminders
Causes
Alarming levels
TEMP Measurement
out of scope
TEMP measurement value out of
measurement scope
High
15.5
Maintenance & Cleaning
Warning
Users must turn off the equipment and shut down the AC power supply before cleaning the
monitor or the connected sensor.
This monitor is compatible with YSI400 series TEMP detectors, whose cleaning procedures are as follows:
Reusable TEMP detectors:
Heating onto a TEMP detector can not be over 100
o
C (212
.
F), as such detector can only undertake
80
o
C (176
.
F) --100
o
C (212
.
F) within short period.
Detectors can not be vapor disinfected.
Only cleaning agents with alcohol can be used for disinfection.
During use of straight detectors, users should cover them with protective adhesive.
When cleaning detectors, users should use one hand to hold on one end and the other hand to
downward hold wet lint-free cloth to wash detectors towards the connector direction.
Attention
If you are using a one-time TEMP detector, this detector is allowed to be re-disinfected or reused.
Attention
To protect the environment, one-time TEMP detectors should be recycled or properly treated.