Operating Instructions
Using Performance Validation
This section describes the benefits of the Performance Validation System and explains how to set up
the metal detector for PV testing, with information about each of the parameters that need to be
specified.
Regular checking of the metal detector’s sensitivity should be a normal part of any quality assurance
procedure. This is particularly important should it become necessary to demonstrate due diligence to
address customer or legislative requirements.
One approach is to make regular sensitivity checks and keep handwritten records of the results.
To assist in maintaining quality assurance controls the metal detector includes an automatic
Performance Validation System. This can be set up to automatically prompt the operator to perform
quality assurance checks on a regular basis. It then guides the operator through a test sequence
using standard test samples. At the end of the Performance Validation test the results are saved in
memory and can be outputted in a number of ways.
See the
Setting Up PV Testing
To use PV testing the PV Test option must be set to YES in the
Menu > Service > Options
menu.
The
Menus > Set Up > PV Test Set Up
menu will then become available.
A series of options the allow you to specify how the performance validation test operates and these
are described below.
The PV Test Set Up menu is not available if the Reverse Mode option has been set to YES
in the Menus > Service > Options menu.
To set up PV testing, proceed as follows:
1.
Select
Menus > Set Up > PV Test Set Up
to display the
PV Test Set Up
menu options
which are listed and described below.
2.
Use the
Navigation Keys
to select each option in turn press the
OK /
Enter Key
to enable the value to be edited.
3.
Use the
Navigation Keys
to select the required option value and then
press the
OK / Enter
to confirm the selection.
4.
Repeat step 2 and 3 until you have selected and confirmed values for each option.
The PV Test Set Up menu options available are as follows:
■
Test type
The following options are available for setting the test type to be used.
O
Normal
Selecting this option enables the normal test process. The size of the Fe (Ferrite), Nfe
(Non-Ferrite) and S.St (Stainless Steel) test wands being used must first be entered and
menu options are available for these.
When the PV test starts the user will be prompted to pass a sample product pack with
the Fe test wand type attached through the detector where it will be inspected to confirm
that the metal contaminant in the test wand can be detected
102
Summary of Contents for IQ3
Page 2: ......
Page 10: ......
Page 26: ......
Page 43: ...Installation 33 ...
Page 60: ...Installation THIS PAGE IS INTENTIONALLY LEFT BLANK 50 ...
Page 70: ...About the Metal Detector Range THIS PAGE IS INTENTIONALLY LEFT BLANK 60 ...
Page 74: ...Getting Started IQ3 Dual Line Display Menu Map 64 ...
Page 116: ...Operating Instructions THIS PAGE IS INTENTIONALLY LEFT BLANK 106 ...
Page 140: ...Maintenance THIS PAGE IS INTENTIONALLY LEFT BLANK 130 ...