Electromagnetic Compatibility (EMC)
The Lode equipment is intended for use in the electromagnetic environment specified below. It is
the responsibility of the customer or user to ensure that the Lode equipment is used in such an
environment.
Emissions Test
Compliance
Electromagnetic
Environment – Guidance
RF emissions
CISPR11
Group 1
The equipmentr uses RF
energy only for its internal
function. Therefore, its RF
emissions are very low and
are not likely to cause any
interference in nearby
electronic equipment.
RF Emissions
CISPR 11
Class B
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
Flicker emissionsIEC 61000-3-3
Complies
The equipment is suitable for
use in all establishments
including domestic
establishments and those
directly connected to the
public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
The Lode equipment is intended for use in the electromagnetic environment specified below. It is
the responsibility of the customer or user to ensure that the Lode equipment is used in such an
environment.
Immunity Test
IEC 60601-Test
Level
Compliance Level
Electromagnetic
Environment - Guidance
Electrostatic
discharge (ESD)
EN 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient/burst EN
61000-4-4
± 2 kV for power
supply lines
±1 kV for input /
output lines
± 2 kV for power
supply lines
±1 kV for input / output
lines
Mains power should be
that of a typical
commercial or hospital
environment.
Surge
EN 61000-4-5
± 1 kV
differential mode
± 2 kV common
mode
± 1 kV differential
mode
± 2 kV common mode
Mains power should be
that of a typical
commercial or hospital
environment.