Switch on again while keeping the
( )
button pressed. The control by the keyboard is
returned when you select ON in SYSTEM PARAMETERS / KEYBOARD ON/OFF.
!
At the next start-up of the ergometer the keyboard is enabled again.
6.5.3
Test Functions
For service and calibration of the device, a number of diagnostic tools are available in the
software. They are hidden under Test functions.
You need a safety code to enter the TESTFUNCTIONS. They are not accessible for end
users. Contact your supplier in case service of the device is required.
If, accidentally, you get
into to the TEST FUNCTIONS it can cause damage to the ergometer operation and
guarantee will be lost!
6.5.4
P In/Out Analog
The ergometer can be controlled by an external device which has an analog output only (i.e.
some older ECG devices). This can be done with the ANALOG mode.
With P In/out Analog the analog scaling can be adjusted. The following scales are available:
!
1 V analog in corresponds with 100 W workload
!
1 V analog in corresponds with 200 W workload
!
1 V analog in corresponds with 500 W workload
Example of most used scaling:
For the best accuracy it is advised to use this scaling. This restricts the P_in to 500 W and
P_out to 1000 W. Only if you need higher workloads, increase to 200 or 500 watt/volt. If you
are unsure, contact your supplier.
SETTINGS
TESTFUNCTIONS
TESTFUNCTIONS
Security code
SETTINGS
P IN/OUT ANALOG
P IN/OUT ANALOG
1 V = 100 W
Summary of Contents for 925900