Lode BV
Excalibur Sport
54
S= step
the workload can be ad or – and the duration of this hyperbolic
stage can be determined; the load is independent of the RPM.
T= torque
the torque in Nm can be adjusted with a certain duration in this fixed
torque stage; load P = 2*
π
*N*T [W] goes linear with cadence.
L= linear
the workload is coupled with RPM with constant
α
as the adjustable
parameter for the desired duration; load P =
α
* N
2
[W] goes up with
square of cadence.
H= heart rate
the workload is regulated as a function of the adjustable heart rate for a
certain duration in this heart rate controlled step.
P= proportional
the workload is increased linear as a function of time (ramp protocol). It
will take the total duration time to reach the programmed workload.
R= Repeat
repeat steps (up to the maximum workload): you have to enter the
workload step and duration only once.
Recovery
the recovery stage is the last stage of a protocol with a fixed workload and
an unlimited duration.
Programming is transparent. You can mix all above types of variables, except recovery which
is always the last stage, within one protocol.
After the confirmation of the workload during the recovery stage the menu PROGRAM BPM
will appear:
Toggle with the
▼▲
keys for the measurement possibilities:
No BPM
no BPM module available, select this option when the BPM is not installed, or
no automatic blood pressure measurement during this protocol;
Repeat
allows you to program a repeat time in minutes during the exercise stages:
PROGRAM BPM
Measurement
No BPM
esc ok
PROGRAM BPM
repeat time = 03 min
back ok
Program BPM
measurement
repeat
esc ok