only exception is when RPM drops below 25. The workload is removed immediately in that
case, otherwise pedalling becomes too hard. As soon as RPM rises above 30, workload is
applied again with the desired regulation speed. The maximal value can be useful for sport
and/or sprint tests.
6.4
Set mode
Set mode is used to define which parameter can be adjusted during the exercise test in
MANUAL mode with
▼▲
keys. With parameter one of the ergometry variables is meant, like
workload, torque, heart rate and linear constant alpha.
6.4.1
Hyperbolic
Hyperbolic mode means that the workload P can be adjusted in watt. The workload is kept
constant independent of RPM. When one of the
▼▲
keys is kept pressed in the hyperbolic
mode, the workload is adjusted per W until the next decade, after that the workload is
adjusted per 10 W until the next hundred and followed by steps per 100 W. This until the
button is released.
Display during workload adjustment:
6.4.2
HR controlled
Heart rate can be adjusted and the workload is regulated until the target heart rate is reached.
This means when the heart rate increases above the target level, workload is decreased and
the other way around. The step size during adjustment is 1W. The units are beats per minute
indicated as [/m]. Display during heart rate adjustment:
Workload regulation as a function of adjusted heart rate is done on a 30 second interval base,
following the steps below (HR= actual heart rate in bpm, tHR= target heart rate in bpm, P =
workload in [W]):
If
HR>tHR+15
then P
new
=P-P/8
Else if
HR>tHR
then P
new
=P-P/16
SYSTEM PARAMETER
SET MODE
SET MODE
HYPERBOLIC
P = 134 W
P MANUAL
♥
= 85/m
♥
MANUAL