11.2.2 Description of Functions Adjustable on the Bed
The iBoard Basic keypad on the Eleganza 4 bed is controlled by a long press that is confi rmed by a beep.
SAFETY
MONITOR
STATUS
ACTION ON THE
BED
RESULTING FUNCTION
BED SIGNALLING
ON
Pressing ON
Bed Exit Monitoring is recalibrated and runs in
Inner Zone monitoring mode.
If ON is pressed at the nurse station, the Bed
Exit Monitoring will start in the Outer Zone
monitoring mode.
All parameters monitored so far are still moni-
tored.
symbol for ACTIVATED (
ON
) is
displayed next to “
SM
“
“
Bed Exit ON
“ is displayed
(Bed Exit Monitoring is activat-
ed)
Pressing PAUSE
SafetyMonitor (SM) pause starts for the time
defi ned in the system (default 15 minutes).
“
PAUSE
“ is displayed
Pressing OFF
SM turns off
symbol for DEACTIVATED
(
OFF
) is displayed next to “
SM
“
OFF
Pressing ON
(weight on the bed
is OVER 35 kg)
SM functions are turned on according to confi g-
uration from the server except for BEA, which
is always enforced even if it is not in confi gu-
ration. BEA always switches on to Inner Zone
mode automatically and the centre of gravity is
calibrated.
symbol for ACTIVATED (
ON
) is
displayed next to “
SM
“
“
Bed Exit ON
“ is displayed
(Bed Exit Monitoring is activat-
ed)
Pressing ON
(weight on the bed
is LESS THAN 35
kg)
BEA does not start, only monitoring of other SM
parameters starts according to confi guration.
BEA switches to a confi rmed alert status.
symbol for ACTIVATED (
ON
) is
displayed next to “
SM
“
“
Bed Exit OFF
“ is fl ashing con-
tinuously (Bed Exit Monitoring is
deactivated)
display shows fault of the insuf-
fi cient load
Pressing PAUSE
No reaction
symbol for DEACTIVATED
(
OFF
) is displayed next to “
SM
“
Pressing OFF
No reaction
symbol for DEACTIVATED
(
OFF
) is displayed next to “
SM
“
PAUSE
Pressing ON
Bed Exit is recalibrated according to the new
centre of gravity and runs in Inner Zone Bed
Exit monitoring mode. If ON is pressed at the
nurse station, Bed Exit monitoring will start in
the Outer Zone monitoring mode.
All parameters monitored so far are still moni-
tored.
symbol for ACTIVATED (
ON
) is
displayed next to “
SM
“
“
Bed Exit ON
“ is displayed
(Bed Exit Monitoring is activat-
ed)
Pressing PAUSE The running pause is reset to a predefi ned time
(default 15 minutes).
“
PAUSE
“ is displayed
Pressing OFF
SM turns OFF
symbol for DEACTIVATED
(
OFF
) is displayed next to “
SM
“
The patient re-
turns to bed
The pause ends without Inner Zone calibra-
tion. Both SM and BEA run with their original
settings (Autoreturn)
symbol for ACTIVATED (
ON
) is
displayed next to “
SM
“