5 Technical Specifi cations
5.1 Technical Requirements on the Part of the Customer
Internet connectivity
Secure connection*
128kbit/s (upload) per 100 beds
400kbit/s (upload) per 100 beds in the graphical
interface
32kbit/s (upload) per 100 beds with export of
data to the hospital system (HIS)
Minimal server requirements
Dual-core CPU 3GHz, 4GB RAM, 100GB HDD
Server operating system
Linux OS (preferred), Windows server (2016, 2019
and later versions)
Data format for export to the HIS
HL7 v2, xml, webservices
Computer network according to standards
LAN
WLAN
standards
IEEE 802.3
IEEE802.11 (802.11b/g, WPA/WPA2-PSK)
Graphical interface for display at the nurse station
IE11 and higher, Google Chrome
* - A secure connection is not a condition on the part of the customer. However, without a secure connection, LINET® is
unable to provide remote installation support in case of complications.
5.2 Technical Specifi cations
Tag (on the wall)
Dimensions
80 x 40 x 17 mm
Weight
100 g
Material
ABS UL 94-HB
Battery
Battery life
Integrated battery, not replaceable
5 years
Tag location (from the ground)
200 mm (margin: +0 mm, -20 mm)
Minimum spacing between Tags
2200 mm
Integration module
Battery
Battery life
Integrated battery, not replaceable
10 years
Bed Exit monitoring
Minimum volume of Bed Exit Alarm (sound pressure level of the
alarm signal)
60 dBA
Medium volume of Bed Exit Alarm (sound pressure level of the
alarm signal)
63 dBA
Maximum volume of Bed Exit Alarm (sound pressure level of the
alarm signal)
69 dBA
The necessary power interruption time after which the alarm
system is unable to reset the alarm that preceded the power
interruption
10 years
Inner Zone dimensions (Inner Zone Bed Exit monitoring)
50 cm x 45 cm