18/46
D9U001SW0-0101_02
Table 1 - Electromagnetic immunity, telecommunication services according to IEC 61000-4-3
Testing frequency
(MHz)
Band
(MHz)
Service
Modulation
Immunity test
level
V/m
385
380 - 390
TETRA 400
Pulse modulation 18 Hz
27
450
430 - 470
GMRS 460, FRS 460
FM ± 5 kHz deviation 1
kHz sinus wave
28
710
745
780
704 - 787
LTE band 13, 17
Pulse modulation 217 Hz
9
810
870
930
800 - 960
GSM 800/900, TETRA
800, iDEN 820, CDMA
850,
LTE band 5
Pulse modulation 18 Hz
28
1,720
1,845
1,970
1 700 - 1 990
GSM 1800; CDMA
1900; GSM 1900;
DECT; LTE band 1, 3,
4, 25; UMTS
Pulse modulation 217 Hz
28
2,450
2 400 - 2 570
Bluetooth, WLAN,
802.11 b/g/n, RFID
2450,
LTE band 7
Pulse modulation 217 Hz
28
5,240
5,500
5,785
5 100 - 5 800
WLAN 802.11 a/n
Pulse modulation 217 Hz
9
NOTE:
No deviations from EMC standard requirements are applied.
NOTE:
No further measures are known to maintain essential EMC safety.
NOTE:
Beds equipped with an integration module operate in the IEEE 802.11 b/g/n standard (2,400 to 2,483.5 MHz,
DSSS modulation (IEEE 802.11 b), OFDM (IEEE 802.11 g/n) 20MHz bandwidth, EIRP = 0.34 W.