6/46
D9U001SW0-0101_02
1.5 Identifi cation Labels
Integration module identifi cation label
Location: Underside of the integration module
Tag localisation identifi cation label (with Tag serial
number)
Location: Bottom edge of Tag
Localisation receiver identifi cation label
Location: Side of the receiver
1.6 Acoustic Signalling
SOUND
MEANING
CONTINUOUS SOUND
Overheating, weight overload, motor overload, battery
overcurrent, mattress failure, unattached mattress, dis-
connected network cable during Bed Exit monitoring
REPEATED INTERMITTENT SOUND
(0.6s sound / 2.6s silence)
STOP error (all STOP buttons disabled)
(melody: 3 beeps, pause, 2 beeps, longer pause, 3
beeps, pause, 2 beeps)
Bed Exit Alarm
(0.3s sound / 1s silence)
Battery discharge
INTERMITTENT SOUND
(4 tones)
Bed is disconnected from the power supply
(lasting 0.3s)
Confi rmation of the desired function
Function blocking
Backrest has a tilt of 30° or 45°
(lasting 0.5s)
Start of service mode or end of service mode
Keypad error (positioning disabled)
(lasting 0.3s)
Failed operation signalling
Bed Exit monitoring not turned on because there is no
patient on the bed
Bed Exit monitoring not turned on because the bed is
not connected to the power supply
(lasting 3 minutes: 1.1s sound / 1.1s silence)
Brake Signal