Maintenance & Service
Chemical Tubes
Using the LI-6400
19-3
Cleaning The End Cap Threads
It is important that the threads be kept very clean. If dust or other debris ac-
cumulates on them, you will have a difficult time putting the end cap on far
enough to compress the O-ring, and the tube will leak.
After you dump out the old chemicals from a tube, use a stiff brush to clean
the threads of both the end cap and the tube barrel. For especially dirty
threads, soak them in water briefly, then wipe them clean and dry.
Do not lubricate the threads. Lubricant will accumulate dust, and aggravate
the problem. When you reassemble the tube, make sure the threads are clean
and dry. No grease, please.
Figure 19-2. Air mufflers in soda lime
and desiccant tubes.
Replacing the Air Mufflers
There are two air mufflers attached to the air hoses inside the chemi-
cal tubes (Figure 19-2). These mufflers may become clogged, re-
stricting air flow through the tubes and thus reducing the maximum
flow rates. (This is one of several things that can reduce flow rates.
See
To replace the mufflers, remove the bottom cap, empty the chemicals
and then remove the top cap. The old mufflers can then be unscrewed
and replaced with new ones. Do not overtighten the mufflers, as they
can break at the threads very easily. There are air mufflers in the spare
parts kit.
Muffler
Muffler
DESIC
Open
BOT
Do Not
TO
this e
Summary of Contents for LI-6400
Page 1: ...Using the LI 6400 Portable Photosynthesis System ...
Page 15: ...Part I The Basics ...
Page 16: ......
Page 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Page 175: ...Part II Useful Details ...
Page 176: ......
Page 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Page 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Page 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Page 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Page 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Page 289: ...Part III Working With Files ...
Page 290: ......
Page 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Page 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Page 375: ...Part IV Configuration Issues ...
Page 376: ......
Page 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Page 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Page 455: ...Part V Maintenance Troubleshooting ...
Page 456: ......
Page 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Page 593: ...Part VI Programming ...
Page 594: ......
Page 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Page 846: ...Index I 16 Using the LI 6400 ...