Display and control panel AMS 300i
Leuze electronic GmbH + Co. KG
AMS 300
i
45
TNT
35/7-2
4V
Position value submenu
Format
… 8,n,1 / … 8,e,1 / … 8,o,1
Selection of the serial communication data mode. This is
specified with the number of data bits, parity (N=none,
E=even, O=odd) and number of stop bits. For example, '8N1'
thus means 8 data bits, no parity, 1 stop bit.
8,n,1
Output cycle
Value input
Output cycle of data in multiples of the AMS 300
i
measure
-
ment cycle of 1.7ms.
The parameter is only valid if cyclical transmission of the po
-
sition values is selected. Cyclical transmission is selected via
the protocol.
1
Position reso
-
lution
0.01mm / 0.1mm / 1mm / 10mm / free resolution
The measurement value can be displayed in these resolu
-
tions.
The value of the free resolution is determined in the "Position
value" submenu in the "Free resolution value" parameter.
0.1mm
Velocity reso
-
lution
1mm/s / 10mm/s / 100mm/s
1mm/s
Table 8.3:
Position value submenu
Level 3
Level 4
Level 5
Selection/configuration option
Description
Standard
Unit
Metric/Inch
Specifies the units of the measured distances
Metric
Counting di
-
rection
Positive/Negative
Positive: The measurement value begins at 0 and increases
with increasing distance.
Negative: The measurement value begins at 0 and decreases
with increasing distance. Negative distance values may need
to be compensated with an offset or preset.
Positive
Offset
Output value = measurement value + offset
The resolution of the offset value is independent of the se
-
lected "Position resolution" and is entered in mm or inch/100.
The offset value is effective immediately after entry. If the pre
-
set value is activated, this has priority over the offset. Preset
and offset are not offset against each other.
0 mm
Preset
The preset value is accepted by means of teach pulse. The
teach pulse can be applied to a hardware input of the M12
PWR connector. The hardware input must be appropriately
configured. See also configuration of the I/Os.
0 mm
Free resolu
-
tion value
The measurement value can be resolved in increments of 1/
1000 within the 5 … 50000 value range. If e.g. a resolution of
0.875mm per digit is required, the parameter is set to 875.
In the activated interface, the measurement value display
must also be set to "free resolution" ("Position resolution"
parameter).
1000
Table 8.2:
RS 422/RS 232 submenu
Level 3
Level 4
Level 5
Selection/configuration option
Description
Standard