27
Leica VT1000S – Microtome
- The warning signal will cease
automatically after the initial-
ization phase.
- Retighten the loose clamping
screws.
If the clattering sound does
not cease once the clamping
screws have been retight-
ened, do not hesitate to im-
mediately call the Technical
Service. Do not use the instru-
ment when in this condition.
- The visible clamping screws
have become loose during
sectioning.
These symptoms may occur
from time to time and are un-
avoidable, as the clamping
screws which have to be op-
erated by the user cannot be
sealed.
6.
Trouble shooting
Error message
Source of error
Corrective action
- Audible warning signal.
(When operating the instru-
ment for the first time or after
the E-EPROM has been ex-
changed.)
- A clattering sound can be
heard.