14
Instruction manual V1.6 – 11/2007
5.
Operation
5.2
The VT 1000 S controls and their function
Attention: Practise working with the controls without a knife
holder inserted. Only insert the knife holder when you are com-
pletely familiar with all control functions.
Scale setting mm/s
0
0.00
0.5
0.025
1
0.05
2
0.10
3
0.15
4
0.20
5
0.25
6
0.50
7
0.75
8
1.00
9
1.75
10
2.50
-
10-Speed rotating potentiometer with scale
-
Function:
Continuous knife feed adjustment from 0.05
- 2.5 mm/s:
Knife return stroke is performed at constant
speed of 5 mm/s.
The additional locking lever (lever in 12 o’clock
position) prevents the speed setting from be-
ing accidentally changed while sectioning is
in progress.
SPEED
-
Control knob with scale from 0 to 10
-
Function:
Continuous adjustment of knife sectioning
frequency (vibration) from 8 - 100 Hz.
Scale setting
Hz
0
0
0.5
8
1
10
2
20
3
30
4
40
5
50
6
60
7
70
8
80
9
90
10
100
FREQ
F R E Q
Locking lever