25
Leica VT1000S – Microtome
- Audible sound signal.
- Error code
E.01xx is
displayed.
xx - there are several error
codes,
00 - there is only one error
code.
- Audible sound signal.
- Error code
E0.200 is displayed.
- Audible sound signal.
- Error code
E0.300 is displayed.
- Audible sound signal.
- Error code
E0.400 is displayed.
- Audible sound signal.
- Error code
E0.5xx is displayed.
- Audible sound signal.
- Error code
E0.600 is displayed.
6.
Trouble shooting
Error message
Source of error
Corrective action
- Button(s) jammed or defective.
- Locking function /
REV
of
REV/
FORW
-button defective.
- Locking function /
REV
of
REV/
FORW
-button defective.
- Feed mechanism defective.
- Important electronic compo-
nent defective.
- Feed motor defective.
- Light barrier error (forward
feed)
- Light barrier error (section
thickness feed)
- Push the button several times
to unlock;
have defective button re-
placed by the Technical Ser-
vice.
- Switch the instrument off; call
the Technical Service.
- Switch the instrument off; call
the Technical Service.
- Switch the instrument off; call
the Technical Service.
- Switch the instrument off; call
the Technical Service.
- Switch the instrument off; call
the Technical Service.