5.
Operation
-
Adjusting the retraction
-
In programming mode, press
DISP
to display the specimen retraction
menu.
-
The indication 'LO' will be displayed.
-
Press the
-/+
button to set a specimen retraction value between 1
and 999 μm. Or to turn off the retraction by selecting '0'.
-
The selected value will be displayed in the FEED window.
-
Press CLR to quit the menu function.
5.3
Adjusting the amplitude
-
To obtain excellent sectioning results, the amplitude requires adjust-
ment according to the specimen type being sectioned.
To this end:
-
With a 2.5 mm Allen key loosen the clamping screw
(1)
and secure the
eccentric on the bottom with your finger. Selectable amplitude posi-
tions are, from left to right: 0.2 mm; 0.4 mm; 0.6 mm; 0.8 mm; 1 mm.
-
Slide the amplitude clamping screw to the desired amplitude position
and retighten.
When adjusting the amplitude setting, do not remove the clamp-
ing screw, simply loosen it.
The instrument is shipped with the amplitude set to 0.6 mm.
Fig. 18
1