6.
Trouble shooting
Error message
Source of error
Corrective action
- Audible sound signal.
- Error code
E0.700
is displayed
for approx. 2 secs.
- Audible sound signal.
- Error code
E0.8xx
is displayed.
- Audible sound signal.
- Optical signal via red LED.
- Audible sound signal.
- Optical signal via red LED.
- Audible sound signal..
- Software detected severe
hardware fault.
- E-EPROM defective.
- The upper limit of the speci-
men feed has been reached.
- The lower limit of the speci-
men level has been reached
(height adjustment of speci-
men via buffer tray).
- User has tried to select a
specimen thickness via the
+/-
button, which is below the
minimum value (0 μm) or
above the maximum (999 μm).
- Switch the instrument off; call
the Technical Service.
- Instrument can still be used,
though there will be certain
limitations: all values will be
set to default values. New
values (sectioning window,
feed, lowering) cannot be
saved.
- Call the Technical Service.
- Leave the upper limit position
(Switch the
UP/DOWN
-button
in
DOWN
direction).
- Mount a new specimen onto
the specimen holder and start
again.
- After unlocking the
DOWN
position the buffer tray is au-
tomatically raised until the
audible and optical signals
turn off.
- Release the
+/-
button.