What to do if...?
73
Leica M525 OH4 / Ref. 10 714 367 / Version -
What to do if...?
Fault
Cause
Remedy
Automatic balancing cannot
be performed.
Zoom cannot be adjusted electrically.
No XY movements are possible at one of
the two CAN handles.
The microscope has not been balanced
exactly in the B-axis.
The push-button for automatic balancing
is flashing, but the acoustic signal is not
sounding (nothing is happening).
Swing arm cannot be moved.
The stand of the Leica M525 OH4 moves.
The range of movement of the
Leica M525 OH4 is limited
(swing, tilt, rotate, XY movement).
Leica M525 OH4 is not correctly
balanced.
Microscope is tilted at too great an
angle.
Failure of the zoom motor.
No XY movements have been set in
the control unit for the CAN handles.
Installed accessory was not turned
back to the working position when
balancing the B axis.
Balancing process still not complete.
Swing arm locked in position.
Footbrake not applied.
Cable laid too tightly.
Video camera was not correctly
mounted and touches the swing arm.
Position of accessory was changed
after balancing.
➩
Align A/B axes on microscope
in parallel (see page 30).
➩
Repeat automatic balancing
procedure.
➩
Press the Zoom rotary button.
➩
Set the zoom by turning the button
(see page 41).
➩
Set the joystick to XY movement
(see page 50).
➩
Balance the B-axis again.
Make sure that the accessory is turned
back to the working position when
balancing the B-axis (see page 28).
Perform intraoperative B/C balancing
(see page 30).
➩
Rotate the microscope to the B-position
and press the Autoblance push-button.
➩
Release the lock (see page 22).
➩
Lock the footbrake in position (see page 35).
➩
Re-lay the cable (see assembly instructions
Leica M525 OH4).
➩
Mount the video camera correctly.
➩
Balance the Leica M525 OH4
(see page 27).
➩
Perform intraoperative AC/BC balancing
of the Leica M525 OH4 (see page 30).
Summary of Contents for OH4
Page 2: ......
Page 12: ...Releases Leica OH4 Service manual Releases 04 2007 Edition 1...
Page 14: ...Technical description Service manual Leica OH4 09 2006 3243 GOR 2 II Doc Code 10 665 742...
Page 26: ...Technical Description Service manual Leica OH4 9 2006 3243 GOR 1 12 Doc Code 10 665 744...
Page 28: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 02 II Doc Code 10665744...
Page 29: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 1 04 2007 Gahe...
Page 30: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 2 Doc Code 10665744...
Page 31: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 3 04 2007 Gahe...
Page 32: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 4 Doc Code 10665744...
Page 33: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 5 04 2007 Gahe...
Page 34: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 6 Doc Code 10665744...
Page 36: ......
Page 38: ...Service Programs Service manual Leica OH4 11 2006 3243 GOR Doc Code 10 665 744 3 II...
Page 96: ...Maintenance service Service manual Leica OH4 stand 04 2007 3243 GOR 6 2 Doc Code 10 665 744...
Page 98: ...Inspection Service manual Leica OH4 stand 04 2007 3243 GOR 7 II Doc Code 10 665 744...
Page 102: ...Inspection Service manual Leica OH4 stand 04 2007 3243 GOR 7 4 Doc Code 10 665 744...
Page 141: ...Leica M525 OH4 User manual 10 714 367 version...
Page 223: ...Technical data 81 Leica M525 OH4 Ref 10 714 367 Version 1200 max 1925 720 x 720 2685 1945...
Page 224: ...Technical data 82 Leica M525 OH4 Ref 10 714 367 Version 1180 1785 430...
Page 225: ...Technical data 83 Leica M525 OH4 Ref 10 714 367 Version...
Page 226: ...Notes 84 Leica M525 OH4 Ref 10 714 367 Version...
Page 227: ...Notes 85 Leica M525 OH4 Ref 10 714 367 Version...
Page 229: ...Leica M525 OH4 Assembly instructions 10 714 368 Version A...
Page 230: ...Leica M525 OH4 Ref 10 714 368 Version A...