Preparation for operation
30
Leica M525 OH4 / Ref. 10 714 367 / Version -
Intraoperative balancing
Warning 4
Danger of injury due to movement of the microscope
during the balancing process.
➩
Do not stand next to the microscope during the
balancing process.
Warning 5
Danger of injury due to swinging down of the
Leica M525 OH4 surgical microscope.
➩
Never balance or re-equip the instrument over the
field of operation.
➩
Balance the Leica M525 OH4 surgical microscope
after re-equipping it.
➩
Do not release the brakes when the instrument is in an
unbalanced state.
Intraoperative AC/BC balancing allows re-balancing during the
operation. Re-balancing may be necessary if, for instance, an
assistant's tube is adjusted.
Either axes A and C or B and C can be balanced.
For best results, it is recommended to balance axes A
and C before balancing axes B and C.
A/C balancing
If work is performed only in the horizontal microscope position,
re-balancing of the A/C axes may be sufficient.
➩
Press the "All Brakes" button on the CAN handle and move
the optical carrier into the A-position.
The mark (2) must be pointing towards A.
1
➩
Press the start button for automatic AC/BC balancing on the
swing arm.
During the balancing procedure, the push-button (2) on the con-
trol unit flashes green and an acoustic signal sounds.
The A-axis is balanced first, followed by the C-axis.
B/C balancing
If work is performed only in the upright microscope position, re-
balancing of the B/C axes may be sufficient.
➩
Press the "All Brakes" button on the CAN handle and move
the optical carrier into the B-position.
The mark (3) must be pointing towards B.
3
2
Summary of Contents for OH4
Page 2: ......
Page 12: ...Releases Leica OH4 Service manual Releases 04 2007 Edition 1...
Page 14: ...Technical description Service manual Leica OH4 09 2006 3243 GOR 2 II Doc Code 10 665 742...
Page 26: ...Technical Description Service manual Leica OH4 9 2006 3243 GOR 1 12 Doc Code 10 665 744...
Page 28: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 02 II Doc Code 10665744...
Page 29: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 1 04 2007 Gahe...
Page 30: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 2 Doc Code 10665744...
Page 31: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 3 04 2007 Gahe...
Page 32: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 4 Doc Code 10665744...
Page 33: ...Service manual Leica M525 OH4 Wiring Doc Code 10665744 2 5 04 2007 Gahe...
Page 34: ...Wiring Service manual Leica M525 OH4 04 2007 Gahe 2 6 Doc Code 10665744...
Page 36: ......
Page 38: ...Service Programs Service manual Leica OH4 11 2006 3243 GOR Doc Code 10 665 744 3 II...
Page 96: ...Maintenance service Service manual Leica OH4 stand 04 2007 3243 GOR 6 2 Doc Code 10 665 744...
Page 98: ...Inspection Service manual Leica OH4 stand 04 2007 3243 GOR 7 II Doc Code 10 665 744...
Page 102: ...Inspection Service manual Leica OH4 stand 04 2007 3243 GOR 7 4 Doc Code 10 665 744...
Page 141: ...Leica M525 OH4 User manual 10 714 367 version...
Page 223: ...Technical data 81 Leica M525 OH4 Ref 10 714 367 Version 1200 max 1925 720 x 720 2685 1945...
Page 224: ...Technical data 82 Leica M525 OH4 Ref 10 714 367 Version 1180 1785 430...
Page 225: ...Technical data 83 Leica M525 OH4 Ref 10 714 367 Version...
Page 226: ...Notes 84 Leica M525 OH4 Ref 10 714 367 Version...
Page 227: ...Notes 85 Leica M525 OH4 Ref 10 714 367 Version...
Page 229: ...Leica M525 OH4 Assembly instructions 10 714 368 Version A...
Page 230: ...Leica M525 OH4 Ref 10 714 368 Version A...